CALIBRATION (Cont’d) 

 
 
 
 
 
 
 
 
 
 

SETTING CALIBRATION 
NOTE:

  Ensure that nothing is in contact with the scale system during this procedure.  Remove 

hands from the system when noting the displayed calibration results.   
 

STEP  1:

    Remove  the  four  (4)  screws  on  the  bottom  of  the  display  housing.    Lift  the  display 

panel.  Put the scale system into the Calibration Mode by switching the calibration switch on the 
display board (Figure 10).  “

CALIBRATION

” will flash on the display.   

STEP 2:

  

Select known calibrated weights, traceable to NIST.   

NOTE:

    The  weights  should  total  at  least  227  kilograms  (half  of  the  maximum  scale  capacity).  

Put aside a subset of weight that is less than half of the combined weight.  

DO NOT USE

 barbells 

or uncalibrated weights. 

STEP 3:

  Press the 

MENU

 button until “

FULL = 453.59 Kg

” is displayed.  Change “

FULL”

 to the 

value of the total weight of the calibrated weights.  Use the 

UP

 arrow button to select the digit and the 

DOWN

 arrow button to change digit.  Press 

ENTER

 to save changes.   

STEP 4:

  Press the 

MENU

 button until “

HALF = 226.79 Kg

” is displayed.  Change “

HALF”

 to 

the value of the subset of the selected calibrated weights.  Use the 

UP

 arrow button to select the 

digit and the 

DOWN

 arrow button to change the digit. 

 Press 

ENTER

 to save changes.   

STEP 5:

  Press 

MENU

 button until “

3 Pt Calibration

” is displayed.  Press the 

UP

 arrow button.   

STEP 6:  

Zero the scale by removing all weight from the platform.  Press the 

UP 

arrow button.   

STEP 7:

  Place the 

HALF

 weight from Step 4 on platform.  Allow weight to stabilize.  Press 

UP

 

arrow to save change. 

STEP 8:

  Place the 

FULL

 weight from Step 3 on platform.  Allow weight to stabilize.  Press 

UP 

arrow to save change. 

STEP 9

:  Press 

ENTER

 to save the calibration, or 

WEIGH 

to exit without saving

.  

 

STEP 10:

  Remove weight and switch the scale system out of the Calibration Mode on the display 

board (Figure 10).   

STEP 11:

  Place the display panel into the display housing.  Align the display panel with the four 

holes in the display housing.  Secure display panel with the four (4) mounting screws. 

CAUTION 

The  integrated  circuits  and  semiconductors  on 
the  printed  circuit  boards  may  be  damaged 
by  electrostatic  discharge  (ESD).    Be  sure  to 
use proper handling precautions at all times.  

Summary of Contents for SR Scales SR555i-KG

Page 1: ...CROBAN_201001 Page 1 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Page 2: ...ECTION 8 STORAGE AND TRANSPORTATION 8 CLEANING AND DISINFECTING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BASIC SYSTEM OPERATION 12 BATTERY REPLACEMENT 14 THEORY OF OPERATION 15 CALIBRATION 16 CONFIGURI...

Page 3: ...AST 1 ea DISPLAY UNIT WITH CABLE FIVE 5 SET SCREWS IN PLACE 1 ea PACKAGE OF SIX 6 C CELL BATTERIES 1 ea MEDICAL GRADE AC POWER SUPPLY SR555i KG AC ONLY 1 ea LOCTITE 1 ea 1 8 ALLEN WRENCH 1 ea QC INSPE...

Page 4: ...into Base Assembly collar and seat firmly Re install set screws in Base Assembly collar and tighten securely NOTE If an optional printer has been ordered go to Page 6 INSTALLING OPTIONAL PRINTER befor...

Page 5: ...der the Cable Clamp and reinstall the Cable Clamp leaving just enough slack to alleviate any tension on the Cable exiting the Platform Securely tighten the Cable Clamp screw Excess cable MUST NOT cont...

Page 6: ...ach handrail Tighten all six 6 screws securely INSTALLING OPTIONAL HEIGHT BAR STEP 1 Figure 4 Remove the set screw on the back of the Display Unit 1 and in the Base Assembly 2 STEP 2 Slide the Height...

Page 7: ...ensure an accurate representation of the patient s weight The low power microprocessor circuitry allows the SR555i KG to derive its power from six 6 common C cell batteries that will provide up to 10...

Page 8: ...n added to help reduce the growth of bacteria as part of a medical center s approach to creating a cleaner healthcare environment For more information visit https www microban com sr scales STORAGE an...

Page 9: ...t areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available d...

Page 10: ...ximately 30 seconds adjustable to 300 seconds ENTER Stores displayed reading in memory RECALL Recalls last stored stable weight height and BMI AVERAGING Automatic digital filter POWER SUPPLY Six 6 C c...

Page 11: ...ains solid when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on dis...

Page 12: ...PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digit when setting time and date DOWN Press...

Page 13: ...ll be using a cane for support on the scale place the cane on the scale while zeroing the system This will ensure that the patient s NET weight will be displayed CALCULATING BMI Press to calculate BMI...

Page 14: ...igure 8 Place the scale on its side and open the battery compartment cover located in the bottom of the scale base STEP 3 Remove and replace ALL six 6 C cell batteries Refer to diagram on the battery...

Page 15: ...meters serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloa...

Page 16: ...e the subset of calibrated weight on the scale Wait for scale to stabilize note scale reading STEP 4 Place all of the calibrated weight on scale Wait for scale to stabilize note scale reading Remove w...

Page 17: ...eights Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 226 79 Kg is displayed Change HALF to t...

Page 18: ...ystem out of the Calibration Mode on the display board Figure 10 STEP 6 Place the display panel into the display housing Align the display panel with the four holes in the display housing Secure displ...

Page 19: ...ttery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label i...

Page 20: ...or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles I...

Page 21: ...g and Service Manual Part No MAN555i KG_MICROBAN_201001 Page 21 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrument...

Page 22: ...MAN555i KG_MICROBAN_201001 Page 22 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstru...

Reviews: