CALIBRATION (Cont’d)
SETTING CALIBRATION
NOTE:
Ensure that nothing is in contact with the scale system during this procedure. Remove
hands from the system when noting the displayed calibration results.
STEP 1:
Remove the four (4) screws on the bottom of the display housing. Lift the display
panel. Put the scale system into the Calibration Mode by switching the calibration switch on the
display board (Figure 10). “
CALIBRATION
” will flash on the display.
STEP 2:
Select known calibrated weights, traceable to NIST.
NOTE:
The weights should total at least 227 kilograms (half of the maximum scale capacity).
Put aside a subset of weight that is less than half of the combined weight.
DO NOT USE
barbells
or uncalibrated weights.
STEP 3:
Press the
MENU
button until “
FULL = 453.59 Kg
” is displayed. Change “
FULL”
to the
value of the total weight of the calibrated weights. Use the
UP
arrow button to select the digit and the
DOWN
arrow button to change digit. Press
ENTER
to save changes.
STEP 4:
Press the
MENU
button until “
HALF = 226.79 Kg
” is displayed. Change “
HALF”
to
the value of the subset of the selected calibrated weights. Use the
UP
arrow button to select the
digit and the
DOWN
arrow button to change the digit.
Press
ENTER
to save changes.
STEP 5:
Press
MENU
button until “
3 Pt Calibration
” is displayed. Press the
UP
arrow button.
STEP 6:
Zero the scale by removing all weight from the platform. Press the
UP
arrow button.
STEP 7:
Place the
HALF
weight from Step 4 on platform. Allow weight to stabilize. Press
UP
arrow to save change.
STEP 8:
Place the
FULL
weight from Step 3 on platform. Allow weight to stabilize. Press
UP
arrow to save change.
STEP 9
: Press
ENTER
to save the calibration, or
WEIGH
to exit without saving
.
STEP 10:
Remove weight and switch the scale system out of the Calibration Mode on the display
board (Figure 10).
STEP 11:
Place the display panel into the display housing. Align the display panel with the four
holes in the display housing. Secure display panel with the four (4) mounting screws.
CAUTION
The integrated circuits and semiconductors on
the printed circuit boards may be damaged
by electrostatic discharge (ESD). Be sure to
use proper handling precautions at all times.