CLEANING and DISINFECTING 

CLEANING 

 

To clean the display / user interface and other scale contact areas: 

 

Use a soft cloth dampened with water and mild detergent to clean scale surfaces. 

 

Wipe surface with clean soft cloth dampened with water and then dry with clean soft cloth.  

 

Do  not  use  abrasive  materials  to  clean  scale  surface  to  prevent  damage  to  the  surface 
finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth. 

 

DISINFECTION 

To disinfect the display / user interface and other scale contact areas:  

 

Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth. Cloth 

cannot  be  dripping  wet.  Follow  manufacturer’s  instruction  on  the  proper  use  of 
commercially available disinfectants. 

 

Disinfectant solutions with 1% sodium hypochlorite or 70% isopropyl alcohol are suitable 
for display / user interface and other scale contact surfaces. 

 

After  disinfecting,  use  a  soft  cloth  dampened  with  clean  water  and  dry  with  a  soft  clean 
cloth to prevent buildup of material on scale finish. 

 

Do  not  use  abrasive  material  to  disinfect  /  clean  scale  surfaces  to  prevent  damage  to  the 
surface finish. 

 

Do not spray liquid directly onto scale surfaces. Use only a damp cloth. 

 

WARNING:

 

DO  NOT  SPRAY  CLEANING  SOLUTION  OR  LIQUIDS  DIRECTLY 
ON SURFACES TO BE CLEANED

 

WARNING:  EXPOSURE  TO  EXCESSIVE  LIQUID  WILL  DAMAGE  USER 

INTERFACE KEYPAD. 

 

WARNING:

 

DO  NOT 

USE  PRESSURIZED  WATER  OR  STEAM.  THE  SCALE 

SYSTEM CONTAINS ELECTRONIC COMPONENTS THAT MAY BE 

ADVERSELY  AFFECTED  BY  EXPOSURE  TO  SUCH  AN 

ENVIRONMENT. 

 

Summary of Contents for SR Scales SR241KG

Page 1: ...ge 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments In...

Page 2: ...ES 2 PACKING CHECKLIST MODEL SR241KG 3 PACKING CHECKLIST MODEL SR241SOKG 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 4 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE 5 CLEANING AND DISINFECTING 6 SP...

Page 3: ...truments Inc PACKING CHECKLIST Model SR241KG Portable Infant Adult Scale PACKING CHECKLIST Model SR241SOKG Portable Scale DESCRIPTION QUANTITY SCALE BASE 1 ea INFANT CRADLE 1 ea 9 VOLT BATTERY 1 ea CE...

Page 4: ...t Cradle slides onto the Scale Base and snaps securely into position REPLACEMENT PARTS and ACCESSORIES SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR241KG SR241SOKG Portable Infant Adul...

Page 5: ...ensures accurate and repeatable weight readings The SR241SOKG Portable Adult Scale is a unique portable scale system that is lightweight and weather resistant It weighs a mere two kilograms This scal...

Page 6: ...se a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfecta...

Page 7: ...CY 0 2 1 digit of displayed resolution over 0 2 kg AUTO ZERO One button operation SR241KG SCALE WEIGHT 3 kg 7 lb SR241SOKG SCALE WEIGHT 2 kg 4 lb AUTO POWER DOWN After 1 minute AVERAGING Automatic dig...

Page 8: ...be displayed ZERO BABY MODE To zero for infant weigh Slide and snap the cradle into place on the scale Place any blankets pads or disposable liners being used for infant s comfort on the cradle before...

Page 9: ...sary attach the baby cradle by sliding and snapping it into place STEP 2 Place the platform on a stable counter or table top STEP 3 Follow the directions for ZERO BABY MODE in the previous section STE...

Page 10: ...2020 SInstruments Inc BATTERY REPLACEMENT REPLACING BATTERIES The display will read LobAt when battery is low and needs to be changed STEP 1 Figure 3 Turn the scale over and locate the battery cover...

Page 11: ...ty voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded system The output of the operational...

Page 12: ...including the caregiver is touching the patient or scale Set scale on smooth hard floor surface Scale reads 999 9 Return to factory for repair For additional information or assistance phone our Servic...

Page 13: ...rranties expressed or implied and of all other obligations or liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special dama...

Page 14: ...No MAN241KG_201006 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc...

Reviews: