SPHINX
Myth 1 Service Manual
11
8. PROBLEMS AND SOLUTIONS
At the moment of writing the Myth 3 has no known
specific problems.
If in the future you encounter any problem(s) you
may enter the info in this table. This table can then
be used for future reference.
Please also send (by fax or e-mail) the specific
information to the
Sphinx International Service
Department
(see page 3): this info can then be
added to the general database to aid others.
Problem
Cause
Solution
Refer to
page…
Bandwidth
≈
100 kHz
Value of C2L, C2R too large
(220 pF)
Change C2L, C2R value to 47
pF
16/17