Sperry instruments DSA2009TRMS Operating Instructions Manual Download Page 7

7

WARNING

WARNING

6. FREQUENCY MEASUREMENT  Hz (See Figure 8)

(1)  Insert test leads 

- Insert the black test lead into the 

COM input and the red test lead into the “Hz ” 

input.  

(2)   Set 

Hz

 function

 - Turn the dial to the 

Hz function. 

(3)  Test & Measure 

 - Place the red and black test leads at both the positive and negative points to be 

measured. The clamp meter will automatically select the proper range and the display will show the True 

RMS value of the measurement being taken while the slave display shows the frequency value of the AC 

voltage. 

                              The maximum input range shall not exceed 30Vrms when measuring frequency.

7. DC CURRENT MEASUREMENT  

A

  (See Figure 9)

(1)   Set  A function

 - Turn the dial to the 

A

 function. 

(2)  Function ZERO Reset 

 - The clamp meter will sense the geomagnetism and magnetic field around it and 

may show a range of numbers before testing. Before test & measuring, be sure to reset the DC current 

function to 0 by pressing the 

ZERO key to clear the clamp meter. 

(3)  Test & Measure 

 - Using the trigger, open the clamp meters jaw and clamp around the conductor. Make 

sure that the conductor is positioned in the center and that the flow of current matches the arrow located 

between the jaws on the body of the clamp meter. The clamp meter will automatically select the proper 

range and the display will show the value of the measurement being taken. The display will also show a 

positive or negative measurement depending on the direction of the current.  

                              The maximum measured current should not exceed 2000A DC when measuring DC 

current.

8. AC CURRENT MEASUREMENT 

Ã

  (See Figure 10)

(1)  Set à function 

- Turn the dial to the 

àfunction.

(2) Test & Measure 

 - Using the trigger, open the clamp meters jaw and clamp around the conductor. Make 

sure that the conductor is positioned in the center and that the flow of current matches the arrow located 

between the jaws on the body of the clamp meter. The clamp meter will automatically select the proper range 

and the display will show the True RMS value of the measurement being taken, while the slave display shows 

the frequency value of the AC current. 
Note: When measuring AC current below 500Hz, set the dial to ˜V + Lo. This is the low pass filter, it will filter 

high-frequency interference caused by frequencies above 7.5KHz to ensure a stable reading.

                              The maximum measured current should not exceed 2000A AC when measuring AC 

current.

9. CURRENT SIGNAL OUTPUT FUNCTION  (See Figure 11)

 

The clamp meter is designed with a signal output function. The current signal measured by the clamp 
meter can be converted into a voltage signal with the ratio of 1A/1mV. This can be measured through the 
output terminal where the user can observe the wave form of the current signal by connecting the output

signal to an oscilloscope.

                              DO NOT connect more than a 5V connection to the output terminal. 

10. SELECT KEY    

The function 

SELECT key is used to switch between functions at the 

Ω

 , 

 measurement settings.

11. MAX/MIN   
Press the 

MAX/MIN key to obtain the maximum or minimum value. After pressing the key, the maximum value hold function is 

activated, the screen will hold the current maximum measured value.
Press the 

MAX/MIN key again and the minimum value hold function will be activated and the screen will hold the current minimum 

measured value. Both functions can be switched by pressing the key. Press and hold the 

MAX/MIN key for 2 seconds disable 

MAX/MIN measurement mode.

Figure 8

Figure 9

WARNING

Figure 10

WARNING

Figure 11

Summary of Contents for DSA2009TRMS

Page 1: ...mpt AC or DC Alternating current or direct current AC Alternating current Danger High voltage DC Direct current Comply with EU standard Continuity Test I DISPLAY SYMBOLS See Figure 2 1 Low Battery 2 M...

Page 2: ...ons Understand and follow all the safety instructions contained in the manual It is essential that all safety instructions are adhered to Failure to follow the safety instructions may cause injury ins...

Page 3: ...nments can endure greater momentary energy than one designed for CAT II CAT I Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device CAT II Primary...

Page 4: ...libration time Ambient temperature 23 C 5 C Ambient humidity 80 RH Temperature coefficient 0 1 accuracy C 1 DC voltage V Input impedance 10M 2 AC voltage V Main display true RMS voltage Main display f...

Page 5: ...the display will show the DC voltage value being measured If the potential at the red test leads is higher than the potential at the black test leads the display will either show a positive voltage va...

Page 6: ...measurement can be achieved by separating the component from the circuit being tested 4 DIODE MEASUREMENT See Figure 6 1 Insert test leads Insert the black test lead into the COM input and the red tes...

Page 7: ...Test Measure Using the trigger open the clamp meters jaw and clamp around the conductor Make sure that the conductor is positioned in the center and that the flow of current matches the arrow located...

Page 8: ...he main display will show Dn Df Df is displayed on slave display screen Df relative value is the last measured value prior to pressing the REL key while Dn is the current measured value After pressing...

Page 9: ...shall not have been altered by the distributor The distributor shall be fully responsible for any warranties the distributor makes to its purchasers which are broader or more extensive than Sperry s l...

Reviews: