background image

 

> > > VISION SCREENER 

 
 

 
 
 

 

 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

Titmus V4 / Titmus V2 

 

 
 
 
 
 
 
 
User Instruction Manual 

 

Summary of Contents for Titmus V2

Page 1: ...VISION SCREENER Titmus V4 Titmus V2 User Instruction Manual...

Page 2: ...me and experienced users of the Titmus V4 Titmus V2 models of the vision screener To ensure the safe operation of the Titmus V4 Titmus V2 Vision Screener the user should read the entire instruction ma...

Page 3: ...Lever 12 Intermediate Vision Testing Using Plus Lens 12 Operating the Titmus V4 using the Control Panel 13 Operating the Titmus V2 14 Testing using the Vision Screener 15 Preparation for Testing 15 Pa...

Page 4: ...user of the presence of important operating or maintenance instructions in the documents accompanying the instrument This instrument is classified as a Type B Equipment Electrical Hazard Prior to any...

Page 5: ...llowing UL CSA and CE Standards USA AAMI ES 60601 1 Canada CSA C22 2 No 601 1 M90 Category 245 Device Type per the Canadian Medical Device Regulations Class I Device CB Scheme IEC 60601 1 Europe EN 60...

Page 6: ...r Often referred to as the TITMUS the older OV7 model as well as the later generation TITMUS Model 2 are still in use in many industrial locations optical practices as well as schools nationwide The s...

Page 7: ...cessory Case with 10 Lens Cleaning Wipes 3 Fog Eliminator Cloths 1 Dust Cover for instrument storage Other standard accessories based on Slide Set Occupational Job Standards Book Aeromedical 32 Lens 2...

Page 8: ...al soft sided carrying case If supplied in a box remove top foam and lift instrument out of the box by holding the sides If supplied in a carrying case unzip the case to remove the instrument Step 1 S...

Page 9: ...s lens H Base of instrument I Base unlock clip to move unit into open position J Plug in for power cord K On off switch L Fuse holder M Plug in for control panel cord Control Panel N Light source cond...

Page 10: ...E Comfortable foam headrest F Slots for intermediate plus lens G Wheel to advance reverse test slides H Base of instrument I Membrane switches J Base unlock clip to move unit into open position K Plu...

Page 11: ...o place A clicking sound can be heard when locked After locking turn the lock slightly to ensure that it does not rotate The instrument now CANNOT be raised for height adjustment and will remain in a...

Page 12: ...ront of the instrument Perform test REMOVE lens upon completion of test CAUTION Do not turn the FAR NEAR lever while the lens is in the instrument This may damage the instrument Using Plus Lens The pl...

Page 13: ...is the default illumination level Press button with the image of the moon for night vision testing R Test Slide position Number indicates the Test Number Refer to your Slide Information Brochure for...

Page 14: ...ng tested These buttons activate Occlude Shutters that occlude each eye clicking sound Indicator light ON indicates that the eye is being tested If indicator light is OFF eye is occluded Left button i...

Page 15: ...on Patient Positioning 11 Raise the instrument for comfortable viewing based on patient height Move instrument up down See section Height Adjustment 12 Ensure lever is in FAR position 13 Follow the Sl...

Page 16: ...ng the Test results Please refer to the Slide Information Brochure Record Forms supplied with your instrument Sample forms are shown below Top section Fill with appropriate information about the patie...

Page 17: ...n Unlock Lock the instrument for height adjustment CAUTION Ensure the height adjustment lock is engaged prior to lifting the instrument This will prevent the instrument from swinging open when carryin...

Page 18: ...ay be easily cleaned with a soft cloth dampened in a mild solution of soap and water Test Slides are in a completely enclosed module hence cleaning is rarely required If required slides may be cleaned...

Page 19: ...enter of the drum to the side 4 Pull on the tab towards the gear side of the drum This will cause the slide to pop up from one side 5 Remove and replace with new slide 6 Insert new slide with the arro...

Page 20: ...the fuse holder will disconnect the power if instrument is inadvertently left plugged in Follow the steps in reverse to assemble Steps 1 Use screwdriver to snap open the tab on the power module 2 Use...

Page 21: ...2 is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Guidance...

Page 22: ...7 P1 2 80 MHz to 800 MHz d 2 33 P1 2 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the separation dista...

Page 23: ...of transmitter M Rated maximum output power of transmitter W 150 kHz to 80 MHz d 1 17 P1 2 80 MHz to 800 MHz d 1 17 P1 2 800 MHz to 2 5 GHz d 2 33 P1 2 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1...

Page 24: ...each eye looks directly at a target but when covering one eye the covered eye turns inwards Esotropia crossed eyes a tendency of the eye to turn inward toward the nose when both eyes are open Exophori...

Page 25: ...a wedge shaped piece of glass that can bend rays of light toward its base Progressive lens an eyeglass lens that incorporates both the distance and near vision correction into one lens without a line...

Page 26: ...Titmus V4 Titmus V2 NOTES Page 26...

Page 27: ...Titmus V4 Titmus V2 NOTES Page 27...

Page 28: ...e 28 Titmus V4 Titmus V2 Sperian Protection Optical Inc 690 HP Way Chester Virginia 23836 USA Tel 800 446 1802 1 804 452 5200 Fax 1 804 530 0568 Email info titmus com Web www sperian com 19213 rev 05...