Sper scientific AquaShock 850034K Instruction Manual Download Page 13

13

Set TDS Factor

1.  Press 

SETUP

 to enter the Set up mode.

2.  Press ↑ or ↓ to select the Set TDS Factor screen (see Fig. A).

3.  Press 

SETUP

 to enter the mode, the current value setting  

will flash (see Fig. B). 

4.  Press ↑ once to increase the value. Hold down ↑ 

 

to continually increase the value.

 

Press ↓ once to decrease the value. Hold down ↓ 

 

to continually decrease the value.

    The range may be set from 0.40 to 1.00.

5.  Press 

SETUP

 once to confirm the TDS factor value 

 

and return to Normal Mode.

6.  Press 

EXIT

 once to return to Normal Mode without  

saving changes.

Fig. A

Fig. B

Cond.

 

value at 

25°C

TDS KCl

TDS NaCl

TDS 442

TDS Your  

Material

ppm

Factor ppm

Factor

ppm

Factor ppm

Factor

84 μS

40.38

0.5048 38.04

0.4755 50.50

0.6563

447 μS

225.6

0.5047 215.5

0.4822 300.0

0.6712

1413 μS

744.7

0.5270 702.1

0.4969

1000

0.7078

1500 μS

757.1

0.5047 737.1

0.4914 1050

0.7000

8974 μS

5101

0.5685 4487

0.5000 7608

0.8478

12880 μS 7447

0.5782 7230

0.5613 11,367 0.8825

15000 μS 8759

0.5839 8532

0.5688 13,455 0.8970

80 mS

52,168 0.6521

48,384

0.6048 79,688 0.9961

Summary of Contents for AquaShock 850034K

Page 1: ...AquaShock Water Purity Kit 850034K...

Page 2: ...itted in any form or by any means electronic mechanical or other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system withou...

Page 3: ...ODUCTION 4 FEATURES 5 MATERIALS SUPPLIED 5 OPTIONAL ACCESSORIES 5 LCD DISPLAY 6 POWER SUPPLY 7 KEYPAD 8 SETUP MODE 9 CALIBRATION 14 MEASUREMENT PROCEDURES 17 SPECIFICATIONS 23 WARRANTY 24 TABLE OF CON...

Page 4: ...Meter is fully shockproof floats and features a double injection molded case with an integral protective soft grip outer layer high end antishock LED backlight display and is powered by the latest env...

Page 5: ...Fahrenheit selectable 99 data points memory for each measurement type Calibration point review Internal clock and calendar Maximum minimum and average Hold function Low battery indicator Backlight MA...

Page 6: ...Scale Unit Current measurement scale unit 3 Measuring Results Current measurement value 4 ATC Automatic Temperature Compensation indicator 5 Temperature Scale Current temperature measurement unit 6 T...

Page 7: ...charged More than 90 life remaining More than 85 90 life remaining More than 80 85 life remaining Less than 80 life remaining Probe Life Icon LCD DISPLAY POWER SUPPLY This meter is powered by one rech...

Page 8: ...unction 2nd Function 1 Power Backlight Low Battery Power on off Backlight on off 2 Setup Setting parameters Confirm setting 3 MI Store data 4 MR Recall data 5 Max Min Ave Max Min Ave 6 Range Switch me...

Page 9: ...on the LCD 4 Press or to select the C or F temperature unit 5 Press SETUP to save and return to Normal Mode 6 Press EXIT to return to Normal Mode without saving the selection Note When the temperatur...

Page 10: ...st the month day hour and minute see Fig D E F 6 Press EXIT to return to Normal Mode without saving Note This procedure adjusts the meter s internal clock which is a 24 hour format only The internal c...

Page 11: ...or NO see Fig C 5 When YES is displayed press SETUP to confirm and all parameters will restore to factory settings except the clock and the meter returns to Normal Mode 6 Press EXIT to return to Norm...

Page 12: ...ode 6 Press EXIT to return to Normal Mode without saving change To calculate the temperature compensation coefficient 1 Set temperature coefficient to 0 2 Put conductivity probe into the standard cali...

Page 13: ...e to confirm the TDS factor value and return to Normal Mode 6 Press EXIT once to return to Normal Mode without saving changes Fig A Fig B Cond value at 25 C TDS KCl TDS NaCl TDS 442 TDS Your Material...

Page 14: ...atically select the correct range The measured value is displayed above the line the set calibration value is shown flashing below the line see Fig A 2 Press once to increase the Calibration value Hol...

Page 15: ...alue is shown flashing below the line see Fig A 2 Press once to increase the calibration value Hold down to continually increase the value see Fig B Press once to reduce the Calibration value Hold dow...

Page 16: ...e is a cyclic display the Conductivity scale will display as in Fig B and Fig C The TDS scale will display as in Fig D and Fig E If there are no calibration points set the display will show see Fig F...

Page 17: ...cm scales and will calculate the TDS value in ppm or ppt according to the TDS conversion coefficient that was entered in the TDS Factor set up procedure During measurement and calibration this meter w...

Page 18: ...seconds to turn the meter off Indicator Light Descriptions Meter Status Indicator Status Power off Light off Power on Green light on for 5 seconds then light off Low battery Red light flashing every...

Page 19: ...ve the current measured value and view the data point number see Fig A 2 Up to 99 Memories can be saved for both Conductivity and TDS If the saved memory exceeds 99 the new data will be written over s...

Page 20: ...ng see Fig A and Fig B The date and time are on cyclical display If there is no saved data is displayed see Fig C 3 Press or to view the saved data points Press to view the data points increasing by 1...

Page 21: ...see Fig A The meter will return to Normal mode Hold 1 Press HOLD to hold the value see Fig A 2 Press HOLD to release the hold value Note When in Hold Mode all other functions will be disabled except...

Page 22: ...layed see Fig C 3 Press MAX MIN AVE the minimum recorded value will appear on the LCD see Fig D and Fig E The date and time are on cyclical display If no measurements have been taken is displayed see...

Page 23: ...0 01 S 0 1 S 1 S 0 01 mS 0 1 mS 0 01 ppm 0 1 ppm 1ppm 0 01 ppt 0 1 ppt Accuracy 0 5 C 0 9 F 1 full scale 1 full scale Input Temperature Probe Connector BNC Connector Calibrate 5 Points 1 point in each...

Page 24: ...not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal and voi...

Reviews: