background image

13

1. Press enter to enter the pressure compensation mode. 

    The default value of 1013 hpa flashes on the display.

2. Press     to increase the air pressure value and     to  

    decrease it.  

3. When finished press enter again to return to P4.0. Hold 

 

    power for 2 seconds to return to normal mode.

CO2 Calibration

The meter is calibrated at 400ppm CO2 concentration at the 

factory. We suggest recalibrating at least once a year for 

the best accuracy. This should be done in fresh air (=equals 

400ppm).  Calibrating the meter in higher CO2 concentrations 

will lead to inaccurate readings. Calibration procedure:

1. With the meter off press and release power,       and   

 

    simultaneously to enter the CO2 calibration mode. “CAL  

    will appear in the lower display and the meter begins  

    counting down for 30 seconds and then begins calibration 

 

    to 400ppm.

Summary of Contents for 900030

Page 1: ...1 Bluetooth IAQ Anemometer 900030 Instruction Manual...

Page 2: ...2 TABLE OF CONTENTS Introduction 3 Front Panel Description 4 Measurement Procedures 7 Set Up Calibration 11 Nice Flow App 15 Error Codes 19 Specifications 21 Warranty 24...

Page 3: ...ndroid apps and export a summary the data together for later analysis The app will also superimpose current readings of all parameters on your smart phone photo of the object being measured Simple use...

Page 4: ...eezes the current reading MAX MIN Maximum and minimum reading AVG Multi points timed average Battery Low voltage indicator ppm CO2 unit m s fpm Air velocity unit RH Relative humidity unit o C o F Temp...

Page 5: ...um and Maximum values Hold down again to return to normal mode In average mode press to record data or start timed recording In setup mode press to enter data settings With the meter on press Lmode to...

Page 6: ...display of air velocity RH Relative Humidity or CO2 In average mode press to view recorded and average reading of all parameters In setup mode press to select category or decrease value Press simultan...

Page 7: ...ff has been disabled Display Units When initially powered on the meter displays air velocity in the upper display and air temperature in the lower display Press Umode to cycle upper display readings o...

Page 8: ...turned on will be displayed 3 Press the up arrow to scroll minimum and maximum parameters in the upper display and press the down arrow to scroll parameters in the lower displays 4 Press Mx Mn again f...

Page 9: ...mal mode Timed Average 1 In normal mode press AVG for 2 seconds and the meter enters multi point average mode Press it again to enter Timed Average mode 2 Press enter to start recording The upper disp...

Page 10: ...the meter is fully inserted and locked into the cone indicates that it recognizes the cone The meter will now automatically calculate air volume without reference to any measurements you may have prev...

Page 11: ...following options P1 0 Unit selection P2 0 Area input P3 0 RH offset P4 0 Pressure compensation P1 0 Unit Selection 1 Press enter to enter unit selection mode 2 Press arrow to switch from metric to im...

Page 12: ...set RH value can be adjusted to any known standard 1 Press enter to enter the RH Offset mode 0 0RH flashes on the upper display 2 Press to increase and to decrease the offset value i e the difference...

Page 13: ...is calibrated at 400ppm CO2 concentration at the factory We suggest recalibrating at least once a year for the best accuracy This should be done in fresh air equals 400ppm Calibrating the meter in hig...

Page 14: ...420 will flash in the upper display 3 After 10 minutes the reading will stop flashing calibration is complete and the meter returns to normal Note Make sure the batteries are at full voltage during t...

Page 15: ...up to 5 meters can viewed simultaneously a Device Name b Display Preferences All measurement parameters are initially displayed You may remove any parameter from the App display simply by clicking on...

Page 16: ...essure at your current location b Camera Allows the App to superimpose measurement data on photos taken with your smartphone These will be saved to your phone s camera file for later viewing or inclu...

Page 17: ...17 d Average Press to enter timed average mode Press to enter multi point average mode Press to record Press to stop recording Files are saved automatically...

Page 18: ...a serial number sequence number and date The latter two can be replaced with a file name of your choosing To change the name Check the box to the right of the file then tap rename at the bottom of the...

Page 19: ...re the temperature reading E031 Temperature circuit damaged Return to Sper Scientific for repair RH E04 RH error caused by temperature error See Temperature errors and solutions above E11 RH calibrati...

Page 20: ...r still persists return to Sper Scientific for repair E01 E33 Volatge too low or CO2 module error Check the low battery icon and replace if necessary Recalibrate and if the error persists return to Sp...

Page 21: ...o C to 60o C 0 1o F 0 01o C 0 9o F 0 6o C Relative Humidity 0 to 99 9 0 1 3 at 10 to 90 RH otherwise 5 Dew Point 8 to 140o F 5 to 59 9o C 0 1o Wet Bulb Temperature 100 to 140o F 20 to 59 9o C 0 1o Air...

Page 22: ...mply with part 15 of the FCC Rules Operation is subject to the two following conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received includi...

Page 23: ...23 OPTIONAL ACCESSORIES 900031 Airflow Cone Kit 900032 Extension Rod...

Page 24: ...rranty does not cover batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the thermometer to expose its electronics will void the warranty To o...

Reviews: