CALIBRATION

Before beginning the calibration procedure, locate the two 

calibration solutions and keep them close to the meter. If 

the calibration is not performed quickly, the meter will return 

to test mode and the calibration will not hold. While this 

does not damage the meter, the calibration will have to be 

performed again.
1.  Remove the 0NTU and 100NTU bottles from their 

protective cases. Tighten both caps and shake the bottles 

gently to ensure they are properly mixed.

2.  Clean the outside of both bottles to ensure there are no 

fingerprints on the glass. Fingerprints or dirt on the glass 

will alter the readings, so it is critical to handle the bottles 

by the cap from this point on.

3. Turn the meter on.
4.  Press and hold the 

TEST/CAL

 button until CAL appears 

on the screen. 

5.  Let go of the TEST/CAL button. The screen will display 

0.0, prompting for the 0NTU standard solution. You will 

have 20 seconds to perform the next step or the meter will 

exit calibration mode and return to test mode.

Summary of Contents for 860040

Page 1: ...Turbidity Meter 860040 Instruction Manual...

Page 2: ...0040 displays the turbidity level of aqueous solutions on a large LCD screen This meter can measure numerous different types of aqueous solutions and offers a two point self calibration feature that c...

Page 3: ...use in the field or laboratory Data Hold Function Optional DC adapter for field use Zero adjustment feature for low turbidity solutions MATERIALS SUPPLIED Turbidity meter Calibration standards solutio...

Page 4: ...ton 7 Zero Button 8 Min Max Button 3 FRONT PANEL DESCRIPTION 3 1 Cover of Testing bottle 3 2 Container of Testing bottle 3 3 Display 3 4 Hold Button Esc Button 3 5 TEST CAL Button 3 6 Power Button 3 7...

Page 5: ...ver 2 Remove the battery cover 3 Insert six new AAA batteries ensuring correct polarity 4 Replace the battery cover and reinstall the screws Replace the batteries when the low battery icon blinks on t...

Page 6: ...ten both caps and shake the bottles gently to ensure they are properly mixed 2 Clean the outside of both bottles to ensure there are no fingerprints on the glass Fingerprints or dirt on the glass will...

Page 7: ...d display the number 100 prompting for the 100NTU standard solution Again you will have 20 seconds to perform this step or the meter will exit calibration mode 8 Place the 100NTU bottle into the test...

Page 8: ...the minimum and maximum values of a set of samples This is especially useful in the case where there are slight changes in the turbidity of many samples or there is heavy sediment in a single sample...

Page 9: ...ity systems such as well water tap water or surface water from lakes and rivers This method is effective for any solutions whose expected value is 2 0 NTU The feature allows you to define a custom sta...

Page 10: ...All readings will be offset by the zero value until the meter is recalibrated or zeroed with the 0NTU standard Data Hold The Minimum Maximum functions on the meter are disabled during data hold Also n...

Page 11: ...on alcohol or benzenes on the meter Repairs or services not covered in this manual should be performed by qualified personnel only Please contact Sper Scientific to speak with a technician Periodicall...

Page 12: ...ot cover sample vials batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal an...

Reviews: