Sper scientific 840080 Instruction Manual Download Page 5

the 

REC 

button.  "

SA

" (save) and "

CA

" will appear on the screen. 

Follow the same procedure for the negative pressure calibration point.  By 
pressing the UNIT button, the LCD now displays  "-80.0" and a small "CA".  
Save the calibration point by pressing the 

REC 

button.  "

SA

" and "

CA

" will 

appear on the screen, followed by "End" and "CA".  In 2 seconds, the  
meter returns to normal mode. 

Press the 

Unit 

button to exit the calibration mode. 

   

Note:  If you can not save by pressing the 

REC 

button, check that the  

calibration pressure source is between 75.0 and 85.0; and that you entered the 
correct positive pressure (+) or negative pressure (-). 
 

3-D.  Auto Power Off 

 

This unit has a built-in Auto Power Off function to prolong battery life. The 
meter will power off automatically if no buttons are pressed within 20 minutes.  
To deactivate this feature,  press 

On

 and 

Hold 

simultaneously before power-

ing on the unit. The symbol "

n

" will appear in the middle of the screen, at which 

time you can release the 

Hold 

button. 

 

3-E.  Battery Replacement 

 

When the "BAT" icon appears on the right side of the LCD, open the battery 
cover, install a fresh 9V battery and replace the cover.  Remove the battery 
during prolonged periods of non-use. 
 

3-F.  RS232 Output 

 

The unit features an RS232 output via which can be utilized by user's specific 
application.  Connection procedures: 

 

Connect the optional RS232 cable to the RS232 output port. 

Plug the other end of the RS232 cable into computer's COM 1 or 2 port.  

Install the optional RS232 Software according to the procedures in the 
software package.  Additional  instructions are found in a separate Adobe 
PDF file in the software.  Be sure to set the baud rate to 2400.

 

 

3-G.  Optional Accessories 

 

840027 AC Adaptor

 

840054 RS232 Cable 

840052 Software

 

840089 Rubber Holster

 

840090 Water Resistant Instrument Pouch

 

840094 USB Adaptor Cable 

Summary of Contents for 840080

Page 1: ...MANOMETERS 840080 840081 840082 840083 and 840085 INSTRUCTION MANUAL SPER SCIENTIFIC LTD...

Page 2: ...atic systems and computer peripherals Model 840085 reads gauge pressure and differential pressure Imperial and Metric in 5 units of measure bar psi inH2O mmH2O and mbar All other models have 11 units...

Page 3: ...ferenced to ambient pressure When the manometer is first powered on the entire LCD display flashes for approximately 3 seconds The LCD is divided into two distinct sections One large Primary top scree...

Page 4: ...splays o o o o one at a time from left to right Next the LCD clears and displays a zero setting Pg 6 Fig I 3 C Calibration The meter is factory calibrated Annual recalibration at a qualified meter cal...

Page 5: ...ure press On and Hold simultaneously before power ing on the unit The symbol n will appear in the middle of the screen at which time you can release the Hold button 3 E Battery Replacement When the BA...

Page 6: ...ange Err 2 Measurement pressure is less than minimum range Err 3 While in the DIF function the differential pressure value is larger than maximum display digit Err 4 During Zero Setting make sure the...

Page 7: ...122 F 0 50 C Power One 9 0 volt battery or AC Adaptor 4 SPECIFICATIONS Model 840085 840080 840081 840082 840083 Range Res Range Res Range Res Range Res Range Res PSI 2 0 004 5 0 003 15 0 01 30 0 02 1...

Page 8: ...nt Sper Scientific product will be substituted if available This warranty does not cover probes batteries or damage resulting from accident misuse or abuse of the product In order to obtain warranty s...

Reviews: