background image

4

PB

3.  OPERATING INSTRUCTIONS
3-A.  MEASURING PROCEDURES

 - See section 2, page 3 for button 

locations and descriptions.
• Connect the 

BNC PLUG

 (4) to the

 BNC SOCKET

 (5).

• Slide the 

PROBE HANDLE

 (3) into the 

PROBE HOLDER 

(2) (opt).

• Press the

 POWER

 (7) button to turn the unit on.

•  The default mode is: Auto range, “A” frequency weighting and  

Fast time weighting.

• To change the frequency weighting, press the 

A/C

 (11) button.  

•  The “A” frequency weighting simulates human ear response. For an 

environmental sound measurement, select the “A” weighting.  

•  The “C” weighting approximates a flat response. Often, “C” is used 

to check the noise of machinery where the target sound level is 

already known.

•  For more information about Frequency Weighting Characteristics, see the 

chart on page 8.

•  Point the

 MICROPHONE

 (1) at the sound source.

•  Use the auto range setting (default) or find the appropriate range using the 

RANGE

 (13) button.  

•  The manual selections are: 30 to 80dB, 50 to 100db, or 80 to 130db.  

•  If “- - - - -” is displayed, switch to another range. The sound level is 

displayed in decibels (dB).

•  For general applications, press the 

FAST/SLOW

 (12) button to select 

“Fast”, simulating the human ear’s response time.  

•  The “Slow” setting is generally used to obtain an average of vibrating 

sound levels.

•  Press the 

POWER 

(7) button to turn the unit off.

3-B.  HOLD

•  During the measuring procedure, press the 

HOLD

 (8) button  

to freeze the display.

•  “HOLD” and the last measurement are displayed. 

•  Press the 

HOLD

 (8) button again to exit this function.

3-C  RECORD MAXIMUM / MINIMUM MEASURMENTS

•  Press the 

REC MAX/MIN

 (9) button once to enter the  

recording mode. “REC” and the recorded values are displayed 

 

and continuously updated.

•  Press the 

REC MAX/MIN

 (9) button as needed to view the recorded Max 

and Min values. “REC Max” and the maximum recorded value, or “REC 

Min” and the minimum recorded value

 

are displayed.

Summary of Contents for 840012

Page 1: ...Detachable Probe Sound Meter 840012 Instruction Manual ...

Page 2: ...easuring Procedures 4 3 B Hold 4 3 C Rec Max Min 4 5 3 D Max Hold 5 3 E Signal Output 5 3 F Calibration 5 3 G Precautions 5 3 H Auto Shut Off 6 3 I Battery Replacement 6 3 J RS232 PC Serial Interface 6 4 Specifications 7 5 Frequency Weighting A and C Characteristics 8 6 Warranty 8 ...

Page 3: ...h Manual and Auto ranging scales with 0 1 dB resolution and an accuracy of 1 5 dB Calibrate in compliance with OSHA using Acoustical Calibrator 840031 or equivalent Has a fold out easel back and a tripod screw 2 METER DESCRIPTION 1 Microphone 2 Probe Holder 3 Probe Handle 4 BNC Plug 5 BNC Socket 6 LCD display 7 Power Button 8 Hold Button 9 Rec Max Min 10 Max Hold Button 11 A C Button 12 Fast Slow ...

Page 4: ...nge setting default or find the appropriate range using the RANGE 13 button The manual selections are 30 to 80dB 50 to 100db or 80 to 130db If is displayed switch to another range The sound level is displayed in decibels dB For general applications press the FAST SLOW 12 button to select Fast simulating the human ear s response time The Slow setting is generally used to obtain an average of vibrat...

Page 5: ...ut device such as an analyzer recorder or controller 3 F CALIBRATION The meter s CAL 15 adjustment is located on the side of the unit Use Sper Scientific s 2 Pt Acoustical Calibrator 850016 or equivalent to calibrate the unit in compliance with OSHA Press the POWER 7 button to turn the unit on Turn the acoustical calibrator on and place it onto the meter s MICROPHONE 1 To calibrate the meter to 94...

Page 6: ...al output is a 16 digit data stream that can be adapted to user defined applications A RS232 lead with the following connection is required to link the instrument with the PC serial interface The 16 digit data stream will be displayed in the following format D15D14D13D12D11D10D9D8D7D6D5D4D3D2D1D0 Each digit indicates the following status D0 End word D1 D8 Display reading D1 LSD D8 MSD Example Disp...

Page 7: ...ut Signal AC output AC 0 5 Vrms corresponding to each range step Output impedance 600 ohm Slow S t 500 ms Output Terminals RS232 AC 3 5mm diameter jack Operating Temp 32 to 122 F 0 to 50 C Operating Humidity Less than 80 RH Power Supply One 9V battery heavy duty or alkaline type Power Consumption Approximately DC 6 mA Weight Weight 8 oz 225 g with battery Dimension Meter 10 6 x 2 7 x 1 1 268 x 68 ...

Page 8: ... to expose its electronics will break the waterproof seal and void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 480 948 4448 The defective unit must be accompanied by a description of the problem and your return address Register your product online at www sperwarranty com within 10 days of purchase Revise...

Reviews: