Page 1: ...Carbon Monoxide Meter 800058 ...

Page 2: ...on Location 7 4 Operating Instructions 9 4 1 Power Up 9 4 2 Selecting the Temperature Units 9 4 3 Data Hold Operation 9 4 4 SET ALARM Operation 9 4 5 MAX Operation 10 4 6 Auto Power Off 10 4 7 Backlight On Off 10 4 8 Low Battery Condition 10 4 9 Beeper Off 11 4 10 Battery Replacement 11 4 11 Calibration or Service 12 5 Warranty 12 2 ...

Page 3: ...andard 62 1989 for living areas 35 ppm The average exposure level per U S OSHA workplace standards 50 ppm OSHA enclosed space 8 hour average level 100 ppm OSHA exposure limit 200 ppm Mild headache fatigue nausea and dizziness 400 ppm Headache occur in 2 5 hours to 3 5 hours 800 ppm Dizziness nausea and convulsions Death within 2 to 3 hours U S Department of Labor Occupational Safety Health Adminis...

Page 4: ... ppm A beeper tone Alarm System The higher the concentration of CO the faster the beeper sounds Above 200 ppm the beeper sounds continuously while the frequency of the tone increases with the concentration of CO Above 1000ppm the beeper sounds continuously with the highest frequency of tone The LCD reads OL to indicate an out of range condition Self Test Install the batteries To turn the meter on ...

Page 5: ...e Typical 3 years Correction of the year Sensor Type Electrochemical Carbon Monoxide Gas Sensor Temperature Sensor Thermistor Accuracy at 23 C Range Accuracy 0 1000 ppm 5 reading or 5 ppm whichever is greater 20 C 70 C 0 8 C 4 F 158 F 1 5 F Resolution 1 ppm 0 1 C 0 1 F Response Time t90 in slowly moving air Carbon Monoxide 60 secs Temperature 10 secs Temperature Coefficient 0 1 X specified accurac...

Page 6: ...h CO Alarm Audible alert when CO level rises above programmed limit default is 35 ppm Sample Rate 2 times per second Battery 4 each AA R6 batteries Battery Life 200 hours continuous with alkaline battery Dimension 190x65x45 mm 7 5x2 6x1 7 inches Weight Approx 235 g Accessories Batteries instruction manual carrying case 6 ...

Page 7: ...3 Symbols and Buttons 1 Sensor 2 LCD Display 3 HOLD Button 4 MAX Button 5 ON OFF Button 6 SET Alarm Button 7 Backlight Button 8 C F Temperature Button 9 Battery Cover 7 ...

Page 8: ...city Indication Auto Power Off Indication Maximum Indication Data Hold Indication CO Reading Indication Beeper Off Indication Alarm Level Unit Indication Unit Indication Temperature Reading Indication SET Alarm Alarm Indication 8 ...

Page 9: ...me the meter is turned on the scale setting will be the same as it was when you powered off last time 4 3 Data Hold Operation The present reading may be held on the display by pressing the button When the hold data is no longer needed release the data hold operation by pressing the button again When the meter is in the Data Hold mode the and buttons are disabled 4 4 SET ALARM Operation Press and h...

Page 10: ...to Power Off By default when the meter is powered on it is in the auto power off mode The meter will power itself off after 30 minutes if no key operation is performed To disable the auto power off feature press and hold the button and then power on the meter The auto power off symbol will disappear to indicate that auto power off is disabled 4 7 Backlight On Off The display backlight is switched ...

Page 11: ...ement 1 Back out the screw at the bottom of the battery door and lift the door away from the meter 2 Remove the four AA batteries from the compartment 3 Replace with four new AA batteries observing proper polarity as depicted on the bottom of the battery compartment 4 Close the battery door and tighten the screw to lock it in place 4 AA size batteries Battery cover screw Fig 1 Battery Compartment ...

Page 12: ...y defective unit without charge If your model has since been discontinued an equivalent Sper Scientific product will be substituted if available This warranty does not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage pr...