background image

 

When the counting exceed 59min 59sec, the time 
scale will be changed to hours and minutes and the 
“ 

H

 ” symbol will appear on the display. 

 

The counting can be reset by press and hold 
“ 

TIMER

 ”

 button for 2 sec. 

4.7 Relative Operation: 

When one press the “

REL

” button, the meter will 

memorize the present reading and the difference 
between the new reading and the memorized data 
will be shown on the display. Press the “

REL

” button 

again to exit the Relative operation. 

4.8 MAX/MIN/AVG Operation: 

When one press the       button the meter will 
enter  
the MAX/MIN mode. Under this mode the maximum 
value, minimum value and average value of latest 4 
data is kept in the memory simultaneously and 
updated with every new data. 
When the MAX symbol is display, the Maximum is 
shown on the display. 
Press     again, then the NIN symbol is on the 
display and also the minimum reading. 
Press      again, the AVG symbol is on the display 
and also the average reading. 
Press    again, MAX, MIN and AVG will blink 
together. This means that all these data is updated int 
the memory and the reading is the present 
temperature. 
One may press       to circulate the display mode 
among these options.   

When the meter is under      operation, “

REL

” 

Summary of Contents for 800005

Page 1: ...302 TYPE K J THERMOMETER T ype K and J T her m om et er 800005 800005...

Page 2: ...3 IV Operation Instructions 5 4 1 Power Up 5 4 2 Connection the Thermocouples 5 4 3 selecting the Temperature Scale 5 4 4 selecting the Thermocouple Type 5 4 5 Data Hold Operation 5 4 6 Timer Operati...

Page 3: ...ons Numerical Display 4 digital liquid crystal display Measurement Range K Type 200 1370 328 2498 J Type 200 760 328 1400 Resolution K Type 200 C 800 C 0 1 C 800 C 1370 C 1 C 328 F 1000 F 0 1 F else 1...

Page 4: ...e following tolerance into the accuracy spec 0 01 of reading 0 03 C 0 01 of reading 0 06 F Note The basic accuracy Specification does not include the error of the probe please refer to the probe accur...

Page 5: ...ndicates that the minus temperature is sensed H It indicates that the timer is expressing in hours and minutes C F Centigrade and Fahrenheit indication K J Thermocouple Type Indication HOLD This indic...

Page 6: ...for proper operation 1 K J type temperature sensor connector 2 LCD display 3 ON OFF button 4 HOLD button 5 Timer function control button 6 Relative readout button 7 MAX MIN Average control button 8 C...

Page 7: ...to Celsius 4 4 Selecting the Thermocouple Type By default when the meter is powered on it is K Type One may press and hold C F button and then power on the meter then it will change to J Type 4 5 Dat...

Page 8: ...one press the button the meter will enter the MAX MIN mode Under this mode the maximum value minimum value and average value of latest 4 data is kept in the memory simultaneously and updated with eve...

Page 9: ...to power off One may press and hold HOLD button and then power on the meter and there will be two successive beeps to indicate that auto power off is disabled 4 10 Low Battery Condition When the batte...

Page 10: ...H AVG MAX MIN button N ASC 4EH Exit AVG MAX MIN mode R ASC 52H REL button C ASC 43H button A ASC 41H Inquire all encoded data Send encoded 8 byte Command K Return 4 bytes For example when sends comman...

Page 11: ...l return MAX Command T Equivalent to one pushing on the HOLD button Command M Equivalent to one pushing on the HOLD AVG MAX MIN button and no message is returned Command R Equivalent to one pushing on...

Page 12: ...bit7 1 0 3th BYTE bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0 no use no use no use Time unit no use X1_ X10 minus OL bit0 1 main window value is OL 0 not OL bit1 1 main window value is minus 0 main window...

Page 13: ...12 6th BYTE first two BCD code of sub window value 7th BYTE last two BCD code of sub window value...