Sper scientific 300100 Instruction Manual Download Page 3

- 3 - 

III.  MEASUREMENT PROCEDURES

 

1. Gently pull upwards on the 

ADJUSTABLE VOLUME WHEEL 

and turn 

it in either direction until the desired setting appears in the 

DIGITAL 

DISPLAY 

(see chart below). 

 

DIGITAL DISPLAY Settings

300100 

 

300101 

 

300102 

 

300103 

 

 

 

 

 

 

 

 

 

1.25µl 

 

7.25µl 

 

12.5µl 

 

75µl 

300104  

 

300105 

 

300106 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

125µl 

 

0.75ML 

 

2.5ML 

 

 

2. To lock in your desired volume, push the 

ADJUSTABLE VOLUME 

WHEEL 

back down into the locked position.  The small grooves in the 

bottom of the 

WHEEL 

align with the handle. 

3. Secure a pipette tip to the 

TIP 

of the unit. 

4. Holding the unit vertically, depress the 

PLUNGER 

to the first stop.  Next 

fully immerse the 

TIP 

2 to 4 mm into the solution. 

5. Slowly release the 

PLUNGER 

until it returns to the original position. 

6. Remove the unit from the solution after 1 to 2 seconds. 
7. To discharge the solution from the pipette tip, place the 

TIP 

against the 

inside wall of the receptacle.  Depress the 

PLUNGER 

to the first stop, 

then depress fully to completely blow out the solution from the 

TIP

8. Next move the 

TIP 

up and along the inner wall of the receptacle several 

times. 

9. You may now remove the unit and release the 

PLUNGER

10. Depress the 

TIP EJECTOR 

to remove the pipette tip. 

11. Clean the unit using an autoclave or ultrasonic cleaner as required by 

your application.  The unit may be autoclaved for a maximum of 8 times, 
after which, the unit should be discarded.  Calibration remains stable 
after occasional autoclaving.  Annual recalibration by a professional 
service center is recommended to ensure accuracy. 

 
Notes:  

Highly concentrated solutions may require increased or decreased 

DIGITAL DISPLAY 

settings.  Make adjustments as needed. 

Inserting objects into the top of the tip assembly can damage the pipette 
seal. 