5
3-E. Scale Selection:
Press the
SCALE
button to toggle
though the scales. The ▼ symbol indicates the active scale.
3-F. High/Low Range Limit Alert
: A red light
will flash
when the measurement is outside of the range limit.
•
Setting the High Range Limit Alert
•
With the unit on, press the
FUNC
button as needed until
H.L Y
and the flashing H-Limit icon is displayed.
•
Press the
READ
button and the last digit flashes.
•
Use the ▼ and the ◄ buttons to set the high range limit.
•
Press the
READ
button activate the function.
•
Press the
FUNC
button as needed to return to the
measurement screen.
•
Deactivating the High Range Limit Alert
•
Press the
FUNC
button as needed until
H.L
n
appears.
•
Press the
Read
button to deactivate the function.
•
Press the
FUNC
button as needed to return to the
measurement screen.
•
Setting the Low Range Limit Alert
•
With the unit on, press the
FUNC
button as needed until
L.L Y
and the flashing L-Limit icon is displayed.
•
Press the
READ
button and the last digit flashes.
•
Use the ▼ and the ◄ buttons to set the low range limit.
•
Press the
READ
button activate the function.
•
Press the
FUNC
button as needed to return to the
measurement screen.
•
Deactivating the Low Range Limit Alert
•
Press the
FUNC
button as needed until
L.L
n
appears.
•
Press the
Read
button to deactivate the function.
•
Press the
FUNC
button as needed to return to the
measurement screen.
3-G. Zero Checking:
Perform Zero Checking if the unit has
sustained a shock, is subjected to temperature changes, or
seems inaccurate.
•
Fill the prism window with a standard solution. This should
be the same temperature as the solution to be tested.
•
With the unit on, press the
ZERO
button. A refractive index
between 1.3329~1.3331 indicates that the unit is accurate.
•
If the unit is out of range, a flashing crosshair symbol
will be displayed, indicating that recalibration is required.
•
Press any button to exit this function.