![Spengler Temp'O Laser Instruction Manual Download Page 15](http://html.mh-extra.com/html/spengler/tempo-laser/tempo-laser_instruction-manual_1346561015.webp)
FR
EN
15
representing body temperature.
e) After the selection, press the
SET
button to save the setting parame-
ters and go to the function setting
F-4
.
Note: the surface temperature is the infrared surface temperature of the
ambient/objects, and the body temperature is the result based on dy-
namic offset of ambient temperature and offset of forehead temperature.
4. Function setting F-4: setting of overall offset
a) After the setting button is pressed and held for 3 seconds, the LCD
screen will display
F-1
.
b) Press the setting button for three more times to display
F-4
and en-
able the characters
0.0
to flash.
c) Now the offset value will be increased by 0.1ºC when the
button is
pressed each time.
d) Meanwhile, the offset value will be decreased by 0.1ºC when the
button is pressed each time.
e) After the selection, press the
SET
button to save the setting parame-
ters and go to the function setting
F-5
.
Note: the actual value of temperature measurement is adjusted with the
offset ranging from -3.0ºC to +3.0ºC according to the actual measurement
distance and environment. For example, the measurement result of a
person’s body temperature is 35.9ºC in a special environment, but the
actual body temperature of the person is 36.6ºC. In this case, the correct
temperature value of 36.6ºC can be achieved through adjustment with the
offset set by F-4. (When the offset is negative, the correct result is the
measurement value minus this value. When the offset is positive, the cor-
rect result is the measurement value plus this value. When the offset is
0.0, the correct result is the measurement value.)
5. Function setting F-5: sound switch mode
a) After the setting button is pressed and held for 3 seconds, the LCD
screen will display
F-1
.
b) Press the setting button for four more times to display
F-5
and enable
the character
1
to flash.
c) Now press the
button to display
1
, i.e., switching on the sound func-
tion.
d) Or press the
button to display
0
with the symbol
on the LCD
screen disappearing, i.e., switching off the sound function.
e) After the selection, press the
SET
button to save the setting parameters
and go to the function setting
F-6
.
6. Function setting F-6: Laser guide mode
a) After the setting button is pressed and held for 3 seconds, the LCD
screen will display
F-1
.
b) press the setting button for five more times to display
F-6
and enable
Temp’O Laser
Notice Temp'O Laser_Notice Temp'O Laser 09/11/10 15:28 Page15