calibration should be done during sunny weather in fresh out-
door air that is well-ventilated.
CAUTION: Do not calibrate the meter in the air with an un-
known CO
2
concentration. It will be calibrated as 400ppm,
which can lead to inaccurate measurements.
Manual Calibration
Place the meter in the calibration site. Turn on the meter and
hold down CAL/ESC and Up/DP/WB simultaneously to enter
CO
2
calibration mode. “400ppm” and “CAL” will blink on the
LCD while performing the calibration. Wait about 5 minutes
for the calibration to complete, which is indicated by the blink-
ing stops and LCD returns to normal mode. To abort the cali-
bration, turn off the meter at any time.
To abort the calibration, press RESET for more than 1 sec-
ond.
ABC - Automatic Baseline Calibration
After enabling ABC (see Setup P4.0, above), place the meter
in a well-ventilated area for 7 days to set it to an ambient
level of 400ppm.
RH Calibration
Humidity calculations on the meter can be recalibrated using
33% and 75% salt solutions. Ambient conditions should be at
25°C or 77°F, with stable humidity.
Plug the sensor probe into a 33% salt bottle. Hold down the
CAL/ESC and M
N
X
/AVG/Down buttons until “CAL” is blinking
on the LCD display. In approximately 60 minutes “CAL” will
stop blinking, indicating the end of this phase. To skip the
75% recalibration, press CAL/ESC to exit.
Plug the sensor probe into the 75% salt bottle, then press the
SET. “CAL” will blink on the LCD until the calibration is com-
pleted in approximately 60 minutes.
The 33% calibration can be skipped by pressing Up/DP/WB
or M
N
X
/AVG/Down during the first 5 minutes of the 33% cali-
bration step above.