3. Grasp the tube by the metal bases located at each end. Applying even
pressure, gently rotate the tube a quarter turn until it loosens. The tube may
now be easily removed from its sockets.
4. Install the new tube by reversing the above procedure.
5. If installation of the new tube fails to yield correct lamp operation, check to
make sure the two pins at each end of the tube are properly connected with
the two copper contacts in each socket.
NOTE
: The tubes in all Q-Series lamps are wired in series. Therefore, if one tube
fails the other tube of the same wavelength will not light, even though it may
be in good operating condition.
FILTER ASSEMBLY REPLACEMENT
Remove the two screws holding the filter assembly and slide it off the open end
of the lamp housing. Carefully install the new filter assembly by reversing this
procedure.
ENVIRONMENTAL SPECIFICATIONS
The Spectroline Q-Series lamps are designed to be safe under the following
conditions:
• Indoor use;
• Altitude up to 2,000 m (6,562 ft.);
• Temperature 5°C to 40°C (41°F to 104°F);
• Maximum relative humidity 80% for temperatures up to 31°C (88°F), decreasing
linearly to 50% relative humidity at 40°C (104°F);
• Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
• Installation Category II;
• Pollution Degree 2
4