A-6
MiniCapt® Mobile Operations Manual
Appendix A
: Communications
Input Registers
30216 – 30217:
Time Stamp
Dual-register representation of data packet time stamp. Integer value (type
defined as time_t) representing start date/time in seconds since January 1,
1970.
30218 – 30219:
Sample Time
Dual-register representation of data packet sample interval processed in
units of seconds. Can be integer or float depending on mode selected. The
integer representation is time in seconds multiplied by 100.
30220- 30221:
Data Packet Status
Values are defined in
30222 – 30223:
flowrate
Dual-register representation of the real-time flowrate in liters per minute. Can
be integer or float depending on mode selected. The integer representation
uses scale factor from register 30011.
30224 – 30225:
Volume
Dual-register representation of the real-time volume in liters. Can be integer
or float depending on mode selected. The integer representation uses scale
factor from register 30012.
30226:
Location
Index of location value associated with data packet. Not used for MiniCapt.
30227:
Number of Particle
Channels
Processed as a variable but fixed to a value of 0 for MiniCapt.
30228:
Number of Analog Channels
Processed as a variable but fixed to a value of 3 for MiniCapt.
30229 – 30234:
Analog data
Dual-register representation of the data packet analog value.
Can be integer or float depending on mode selected. The integer value
represents the total averaged Flow, total Volume and total Duration for the
data processed thus far in the recipe. The integer scale factor for Flow is the
same as that described for registers 30222 – 30223. The integer scale factor
for Volume is the same as that described for registers 30224 – 30225. The
integer scale factor for Interval is 1000.
Table A-2
Input register – Data Packet
Input Registers
Description (Data Packet)
Comment
Notes
30201
unary
Calibration Date (Month)
Calibration Date
30202
unary
Calibration Date (Day)
Calibration Date
30203
unary
Calibration Date (Year)
Calibration Date
30204
string
Serial Number: Char 00, 01
Serial Number
30205
string
Serial Number: Char 02, 03
Serial Number
30206
string
Serial Number: Char 04, 05
Serial Number
30207
string
Serial Number: Char 06, 07
Serial Number
30208
string
Serial Number: Char 08, 09
Serial Number
30209
string
Serial Number: Char 10, 11
Serial Number