Charging the Battery
2.5
Charging the Battery
The flame simulator uses lithium-ion batteries as a rechargeable power
source. When the batteries are fully charged, the simulator operates at least
1,000 times without recharging. When the voltage from the batteries is
lower than the required operational level, the simulator will not operate.
To Charge the Battery:
1
Place the flame simulator on a table in a safe area, not exceeding
104°F/40°C.
2
Release the locking screw (Figure 1, Item 8).
3
Unscrew the battery back cover counterclockwise (Figure 1 and Figure 4,
Item 3).
4
Unscrew the locking disc clockwise (Figure 4, Item 4).
5
Pull out the battery from the flame simulator.
6
Connect the battery to the charger. Verify that the charger is supplied
with the flame simulator model FRIWO MPP15, with maximum charging
voltage of 16.8V (4.2V x 4) and maximum current of 700mA.
7
Charge for a maximum of 2–3 hours, until the green LED on the charger
lights.
8
Disconnect the charger.
9
Insert the battery into the flame simulator.
10
Screw on the locking disc (Figure 4, Item 4).
11
Screw on the back cover (Figure 1 and Figure 4, Item 3).
12
Lock the back cover with the locking screw (Figure 1, Item 8).
To Replace the Battery:
1
Place the flame simulator on a table in a safe area, not exceeding
104°F/40°C.
2
Release the locking screw (Figure 1, Item 8).
3
Unscrew the battery back cover counterclockwise (Figure 1 and Figure 4,
Item 3).
4
Unscrew the locking disc clockwise (Figure 4, Item 4).
5
Pull out the battery from the flame simulator.
6
Insert the new battery pack in the simulator housing. Use only a
SPECTREX battery pack, P/N 380004.
7
Screw on the locking disc (Figure 4, Item 4).
8
Screw on the back cover (Figure 1 and Figure 4, Item 3).
9
Lock the back cover with the locking screw (Figure 1, Item 8).