4. Place the black anodized flow-through cell on the main plate, in the
same location as the V-block, with the end closest to the cell
mounting bracket as close as possible to the collector lens
.
Note: the collector lens is the one with a black target in its center.
5. Secure the cell on the main plate with the screws supplied.
6. Start the liquid to be tested flowing through the cell and double check
for any leaks.
7. Preferably set the flow of the liquid to 100 ml/min, with a flow control
downstream
from the cell. This avoids air bubbles forming in the cell.
8. Set the software on the computer connected to the particle counter to
Automode
and follow detailed instructions.
9. Contact Spectrex to obtain Spectracount™ software for Automode
operation to continuously monitor flowing samples over time.