Oxygen Analyzer
4
–
12
4900002239 rev. G 7-11-17
Measurement Control
This register is used to activate interval measurement and also to trigger a
measurement. It is bit-coded as shown in Table 4–3
Table 4-1
Holding registers (Continued)
Register
Name
Address
Size
Variable
Type
Description
Write
Access
LED Intensity
5711
2
Int32
The signal LED intensity.
The allowed range is 0x00
(lowest) to 0x0A (highest)
Yes
Timestamp
8231
2
Int32
This is the current system
time, defined as the num-
ber of seconds that have
elapsed since 00:00:00,
Thursday, 1 January 1970
(Unix time, ISO8601).
NOTE
: Values below
1493050000 will lead to
error code “illegal value”.
Yes
Table 4–2
Definition of register 5707
Start
register
Number of
registers
Reg3 / Reg4
Write
Access
5707
2
Int32:
Bit codes control regis-
ter. Refer to Table 4–3.
Yes
Table 4–3
Measurement control register bit definition
Bit
Description
0
Interval on/off (delete to turn off, set to turn on)
1
Status-Bit: Set when a measurement is currently performed.
Will be deleted after completing the measurement. Setting
this bit does not trigger any action.
2
Start measurement (Single scan or continuous)
3 - 31
Reserved
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...