Oxygen Analyzer
3
–
22
4900002239 rev. G 7-11-17
3.
Press the
Start
button next to the
Cal2nd
box.
The
Status
field will display the message, “Wait - Stabilizing!”. Wait
for the phase values to stabilize (the variation of the phase angle
should be smaller than
.01
until the
Status
field displays the
message, “Ready to Set Value!”
4.
Press the
Start
button to the left of the
Cal2nd
value.
5.
Press .
Once the calibration values have been set:
1.
Press the
Save
button at the upper right of the screen.
2.
Press
to store the calibration data for the selected sensor.
The display will switch to the measurement screen automatically.
Relative Accuracy Test Audit (RATA)
RATA is accessible from the
RATA
button on the SENSOR / SENSOR OPTIONS
menu screen.
1.
Select RATA in the SENSOR OPTIONS window. Refer to Figure 3–27.
Figure 3–27
Sensor Options screen
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...