When considering a location for the installation of the
RO System, consider the following factors:
Light Sources
1. Most of the components of this system are plastic and
are subject to damage by ultraviolet light from the sun
and other sources such as metal halide lighting.
2. Algae is more likely to thrive inside the clear filter housings
when exposed to bright light.
3. Avoid installing this unit in bright light or direct sunlight.
Temperature Extremes
1. The unit must be kept out of areas that are subject to freezing
temperatures.
2. High temperatures greater than 100° F (38° C) must be
avoided. If the unit is used outside, avoid putting the
system in direct sunlight or connecting it to a
garden hose that may be exposed to sunlight.
1. It is recommended that you store your RO System in a cool place
when not being used. (a Flush Valve Kit is recommended for
manually operated systems)
2. If your system is exposed to sunlight you will grow algae in the
housing and it may damage your system. Your RO System must be
protected from freezing or temperatures above 100° F (38°C).
3. MEMBRANE WARNING: All SpectraPure RO membranes, except for
encapsulated membranes, must remain moist at all times. It is the
customer’s responsibility to inspect the membrane upon receipt and
maintain adequate moisture.
Replacement membranes should be kept in the sealed non-permeable
shipping bag and in a refrigerator until use. The membrane can be kept
there for up to 1 year.
(DO NOT FREEZE)
STORAGE:
CHOOSING A MOUNTING LOCATION: