Special Projects Group SM302 User Manual Download Page 10

 

 

3.5

 

Slit changes 

 

The standard slit size of the SM302 is 50

µ

. The slit is located in the back of the housing of the SMA 

flange on the SM302 entrance. If a different slit size is preferred you must order the new slit by contacting 
Spectral Products Spectral Products Group at (505) 296-9541. 
 

The replacement slits are shipped mounted on a paddle assembly. The paddle will fit in the back end 

of the SMA flange. The following are steps necessary to install your new slit: 
 
 

Warning: 

Remove power from the Spectrometer and disconnect cables from the Camera. 

 

1.)

 

Remove the four 4-40 screws from the SM302 hood and remove. 

For the 

Extended

 version you 

must first remove the LPF1300 knob by loosening the bottom set screw.  Please make note of the 
orientation of the knob before removing. 

2.)

 

Remove the two 4-40 screws from the front face of the SMA Flange. Remove flange and existing 
slit paddle. 

3.)

 

Place SM302 on its side with the entrance face up. 

4.)

 

Carefully place the new slit paddle in the slot that is located on the back of the SMA flange. 

5.)

 

Place SMA flange on the SM302 entrance and secure with the two 4-40 screws. 

6.)

 

Reverse process for the hood replacement. 

 

3.6

 

Grating Change 

 

The SM302 must be returned to the factory for graing addition / change. 

 

Spectral Products

 

 

                2659-A Pan American Frwy., NE  Albuquerque, NM 87107   Tel (505) 343-9700    Fax (505) 343-9705 

www.spectralproducts.com 

- 10 -

Please contact Spectral Products Spectral Products Group at (505) 296-9541 for a RGA number. 

Summary of Contents for SM302

Page 1: ...SM302 InGaAs Array Spectrometer User Manual 1105811 August 2006 Spectral Products 2659 A Pan American Frwy NE Albuquerque NM 87107 Tel 505 343 9700 Fax 505 343 9705 www spectralproducts com 1...

Page 2: ...2 2 Attaching the SM302 cables 7 2 2 3 Connecting the Spectrometer power supply 7 3 0 OPERATION 8 3 1 SM302 Spectrometer Serial Port Layout 8 3 2 SM302 Spectrograph Optical Layout 9 3 3 SM302 Spectro...

Page 3: ...m lack of operator attention to proper procedures failure to follow operating instructions unauthorized modifications and natural disasters are not covered under this warranty The SM302 does not conta...

Page 4: ...is thermoelectrically cooled and temperature stabilized to ensure long term operation stability The multichannel detection with a readout rate of 333 kHz or faster both reduces measurement time and e...

Page 5: ...ended Only LPF 1300nm 2500nm InGaAs power supply Input voltage 100 120V AC with DB15 connector SM302 spectrograph power supply UL listed 110 220 V power pack meets or exceeds UL1950 CSA 1402C and IEC...

Page 6: ...2 1 4 Checking System Package Contents Check to make sure that your package contains all of the following components The SM302 system consists of 1 SM302 SM302 EX unit 1 SM302 SM302 EX Power supply N...

Page 7: ...Remove the computer s cover by removing its mounting screws and sliding off the cover 4 Remove the slot bracket by unscrewing the holding screw and sliding it out Save this screw for securing the boa...

Page 8: ...3 TxD Data in from computer to Spectrometer 4 DTR Not used with the Spectrometer 5 GND Signal ground 6 DSR Not used in Spectrometer 7 RTS Request to send from computer 8 CTS Clear to send from Spectro...

Page 9: ...ENTS PER ANGULAR STEP FOR DIFFERENT GRATINGS Lower wavelength scan limit is zero Grating Upper Limit Grv mm nm 1200 1500 600 3000 300 6000 150 12000 75 24000 50 36000 45 40000 3 4 Using the LPF 1300 F...

Page 10: ...sion you must first remove the LPF1300 knob by loosening the bottom set screw Please make note of the orientation of the knob before removing 2 Remove the two 4 40 screws from the front face of the SM...

Page 11: ...02 Power Supply To set power supply to run the SM302 standard jumper to R12 is set for 5V JP1 To set the SM302 Extended jumper R4 is set for 6V JP2 Spectral Products 2659 A Pan American Frwy NE Albuqu...

Page 12: ...NE Albuquerque NM 87107 Tel 505 343 9700 Fax 505 343 9705 www spectralproducts com 12 C Spectrometer Power Supply 1 3 4 5 2 C Grating Selection Pin Dual 1 Common 2 Common 3 5 0V Imin 0 5A Imax 7 0A 4...

Reviews: