SPEC DSDK Quick Start Manual Download Page 2

SPEC Sensors 

DSDK Tool 

www.spec-sensors.com 

Quick Start Guide, Rev. 0.1 

(510) 574-8300 

 

3.2.  Setup 

 

Under the 

Setup

 tab, type the COM port number in the 

COM Port

 field

 

 

Press 

Open COM Port

 

 

3.3.  Configure the D-ULPSM for a Different Gas Sensor 

The DSDK ships from SPEC Sensors pre-configured for the 
CO sensor. To configure the module for another sensor, 
follow the instructions below. Otherwise, skip to Section 3.4.

 

 

Always remove the gas sensors from the module 
before reconfiguring the module. 

 

 

Use a barcode scanner or smartphone app to read and 
decipher the 2D barcode on the back of the sensor.

 

o

 

The sensor barcode includes the following information: Unique sensor serial number, sensor part 
number, target gas, date code, and sensitivity code. For more information, refer to SPEC Sensors 
document 590-002. 

 

 

If necessary, transcribe the deciphered code into the barcode field to the right of the button. Double-check 
for accuracy. Then press 

Set Barcode

.

 

 

The module will take up to 1 minute to reconfigure the settings and calculate parameters. 

 

 

Wait until the module returns: “Set OC….done”, then 
plug the sensor into the module and allow the sensor to 
stabilize. The stabilization time may vary among 
sensors, depending on the target gas, bias settings, 
gain settings, and history of use.       

 

 

3.4.  Read Gas Sensor Measurements 

 

Under the 

Read

 tab, press the 

Single Read PPB

 

button.

 

o

 

The ppb gas measurement will display in the 
field next to the button

 

o

 

The full serial string will appear in the field at 
the bottom of the window

 

o

 

The format of the full serial string is:

 

 

032116030233, 0, 24, 37, 32697, 26604, 22706, 00, 02, 06, 46 

 

10 

11 

032116030233 

24 

37 

32697  26604  22706 

00 

02 

06 

46 

Unique Sensor Serial Number 

Gas Concentration (ppb) 

Temperature (°C) 

Relative Humidity (%) 

Gas Sensor Measurement (ADC counts) 

Temperature Sensor Measurement (ADC counts) 

Relative Humidity Sensor Measurement (ADC counts) 

Timer, Days on power (up to 31 days) 

Timer, Hours on power (up to 23 hours) 

10  Timer, Minutes on power (up to 59 minutes) 
11  Timer, Seconds on power (up to 59 seconds) 

 

 

To start logging consecutive measurements to a file, press the 

Start/Stop Logging

 button.

 

o

 

Approximately once per second, consecutive measurements will be displayed in the field at the 
bottom of the window.

 

o

 

The measurements will also be saved under a default file name in the default file location, indicated 
below the buttons.

 

o

 

If the default file already exists, additional data will be appended to the end of the file. 

 