DSDK Tool 

 

The  DSDK  Tool  is  a  simple  GUI  that  is  useful  for  interacting  with  the  D-ULPSM  component  of  the  DSDK.  The  tool 
allows you to: 

 

(Setup) 

 

Establish serial communication. 

 

Reconfigure the module using the barcode on the back of the SPEC Sensor. 

(Read) 

 

Query the module for a single measurement. 

 

Start & Stop measurement data logging to a file on the computer at 1 s intervals.   

 

Query the module for the contents of the EEPROM memory. 

(Advanced) 

 

Re-zero the gas measurement calibration. 

 

Re-span the gas measurement calibration. 

 

Adjust the temperature sensor offset calibration. 

 

Reconfigure the SPEC Sensor potentiostat control circuit. 
 
 

1.  Required SPEC Sensor Hardware 

 

 

D-ULPSM Rev 0.3 (available individually or as part of DSDK).

 

 

One or more SPEC Sensors (in pinned package).

 

 

USB-to-UART Module with Pinout Adapter Board (included with purchase of D-ULPSM or DSDK).

 

 

2.  Required 3

rd

-Party Hardware 

 

 

Computer with USB port and Windows XP or later. 

  

3.  Instructions 

 
3.1.  Getting Started 

 

Connect D-ULPSM Rev 0.3 module to the USB-to-UART Module via the Pinout Adapter Board.  

 

 

Connect the assembly to a USB port on a Windows PC.  

 

Locate and open the 

Device Manager

 

Expand the 

Ports (COM & LPT)

 heading. 

o

 

If the USB-to-UART Module mounted properly, you will see 

Silicon Labs CP210x USB to UART 

Bridge (COMXX)

, where 

XX

 is replaced by the number of the associated COM port. 

o

 

If it did not mount properly: locate, 

download and install “CP210x USB to UART Bridge VCP 

Drivers

” from the internet.   

 