For slings with rigid supports, remove the plastic inserts before washing. (Look
Cleaning of the lift
Cleaning of the sling
DO NOT USE BLEACH.
REPLACE a sling when it show signs of deterioration.
Deterioration of sling
Disinfecting of the lift and sling
Troubleshooting Guide
The following list of encountered problems and solutions will assist you in
determining what may be causing your patient lift not to function as designed. If you
have a problem occurring which is not listed below please contact your dealer or
technical support for help. Do not attempt to repair or replace components or parts
on your lift as this may void your warranty or cause further problems that may result
in patient injury. Stop using your lift immediately if it is not functioning correctly or any
warning beeps are heard.
I need to re-charge my batteries often or they fail to hold a charge when I
charge them.
•
Replace the batteries or battery pack as they are at the end of their life cycle
• Batteries should be changed every 18-24 months depending on usage
The actuator will either go up or down but not both
•
Replace your hand control as it is at the end of its life cycle
•
Your control box may be malfunctioning
There is a grinding sound inside the actuator when lifting
•
Replace your actuator as the internal gears are stripped
The actuator stops and starts while lifting or lowering
•
Replace the batteries or battery pack as they are at the end of their life cycle
My lift will not operate even though it shows the batteries are charged
•
Unplug the AC power cord from the control box as the lift will not work when the control box/
charger is connected to AC Power.
•
Check to make sure the hand control is properly inserted all the way into the control box
port.
•
Check to make sure the actuator plug is properly connected and inserted all the way into the
control box
•
Your hand control needs to be replaced as it is at the end of its life cycle
If there is a problem you could not solve, please contact your dealer for assistance.