www.spacelabshealthcare.com
4 - 2 9
9 1 3 9 0
Q U B E
M
A I N T E N A N C E
Figure 4-25 Recorder Assembly and CPU PCBA
To replace the recorder
1
If the recorder assembly is installed, open the printer door.
Loosen the two captive screws. Pull out the assembly at the
top. Refer to
2
To reinstall, use the instructions in
Insert Recorder Paper
Two rolls of thermal recorder paper are included with the optional
recorder assembly.
To insert a roll of paper
1
Press the release bar at the right of the recorder assembly.
The front of the recorder assembly is hinged at the bottom.
Two spoon-shaped arms that hold the paper roll spindle
between them drop down.
2
Unroll a short length of paper from the roll. Orient the roll so
that the paper feeds from the bottom.
3
Slip the paper roll spindle between the plastic arms. Close the
front of the assembly so that the end of the paper roll sticks out
of the recorder assembly.
Dock Connector Covers for Option A
The covers are snap-on and snap-off, and can be removed by hand.
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...