4 - 2 3
9 1 3 9 0
Q U B E
M
A I N T E N A N C E
Connector Board PCB
1
Perform
access the Main Chassis assembly.
2
Remove the three screws, shown in
for option A.
Remove the two screws for options B or C.
Figure 4-18 Connector Board PCB
Note:
Loosen the top three Monitor Dock PCB screws.
3
Remove the Connector Board.
4
Install the new Connector Board making sure that the connection
to the interface PCB is good.
Note:
Tighten the top three Monitor Dock PCB screws.
5
Connect board to the Interface PCBA. Connect to Monitor and
Dock boards.
6
Install the three screws removed in step 2 for option A. Install the
two screws removed in step 2 for options B or C.
Remove
Loosen
Connector Board
Remove
Interface PCBA