www.spacelabshealthcare.com
4 - 7
9 1 3 9 0
Q U B E
M
A I N T E N A N C E
An alternate method is to connect the cable and turn the monitor
ON and then OFF. Each color of the embedded alarm light
momentarily illuminates. An external alarm light also briefly
illuminates.
Ethernet (Wired)
1
Attach a 10/100BaseT cable from an active network onto the rear
panel connector.
2
Verify that the green link light on the connector is ON.
Assembly/Disassembly Procedures
Note:
The external power supply is not designed for disassembly.
Required Tools and Parts
•
Anti-static mat with wrist strap
•
#1 and #2 Phillips-head screwdriver
•
3/16-inch nut driver
•
Tool, antenna extraction, wireless card, P/N 003-0286-00
•
XPREZZON™ and qube™ System Administration Manual
, P/N
070-2380-xx
Setup for Disassembly
The battery switch has symbols to show when to turn the switch on
and off. Turn the switch off for shipping and service (refer to
). Turn the switch to on for regular operation (refer to
).
Figure 4-1 Battery Switch Off Symbol
Caution:
Before you begin any disassembly procedures, power OFF the monitor,
disconnect the AC cord from the AC power receptacle, and remove the
batteries.
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...