Figure 4-2 Battery Switch On Symbol
For Option A
1
Remove the external power supply cable and battery or batteries,
if present.
2
Remove the patient parameter module.
For Options B and C
1
Remove the external power supply cable and turn the battery
switch to off.
2
Remove the patient parameter module.
Remove the Bezel Assembly
1
Place the monitor face-down on a stable and padded surface.
2
Remove the eight finishing plugs from the rear enclosure as
shown in
for option A.
3
Remove the 10 screws from the rear enclosure, as shown in
.
Summary of Contents for qube 91390