![Spacelabs 91518 Service Manual Download Page 7](http://html1.mh-extra.com/html/spacelabs/91518/91518_service-manual_1339012007.webp)
91518 Multigas Analyzer Service Manual
1-3
Introduction
Environmental Requirements
lists the environmental requirements, including temperature, for operating and storing the 91518
Multigas Analyzer.
Table 1: Environmental Requirements
Operating
Storage
Temperature
18° to 28° C (64.4° to 82.4° F),
±5° C of calibration
-25° to 70° C (-13° to 158° F)
Humidity
20 to 80% (non-condensing)
relative humidity (RH), ±20%
RH of calibration
10 to 95% (non-condensing)
Altitude
N/A
12,192 m (40,000 feet)
Atmospheric Pressure
500 to 800 mmHg, ±50mmHg
of calibration
N/A