• Enable or disable minor graticules and adjust how they show. Minor graticules
are small, dashed line segments that show between full-scale graticule lines
when scaled pressures show.
To enable or disable the clock display
1
Touch
Monitor Setup
.
2
Choose one of the following according to whether you want to
change the clock setting on the local monitor or a remote
monitor:
• For the setting on the local monitor, touch the
Setup
tab.
• For the setting on the remote monitor, touch the
Remote Config
tab, and
then touch
Remote Setup
on the lower-right corner of the window. The
Remote Setup window shows.
3
Under
Clock
, touch
On
or
Off
.
To display the clock as a stopwatch or timer
1
Make sure the clock display is enabled, as described in
2
Touch the clock display in the lower-right corner of the screen.
3
Touch
STOPWATCH
or
TIMER
.
To return the clock to its digital display, touch the
Clock
in the lower right
corner of the monitor, then select
Digital
from the menu.
To select the number of parameter zones
1
Touch
Monitor Setup
.
2
Touch the
Setup
tab.
3
Under
Zone Format
touch
6
or
8
.
To enable or disable the display alarm lights
1
Touch
Monitor Setup
.
2
Touch the
Setup
tab.
3
Under
Alarm Light
, touch
On
or
Off
.