background image

 

 

3

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

MPK-2000e 

Digital micro-ohmmeter

 

 
 
 

User’s guide 

 
 
 
 
 
 
 
 
GU-1294 
© 2008 SOURCETRONIC. All rights reserved.  
Printed in Germany. 

Summary of Contents for MPK-2000e

Page 1: ...3 MPK 2000e Digital micro ohmmeter User s guide GU 1294 2008 SOURCETRONIC All rights reserved Printed in Germany...

Page 2: ...test leads during the measurement There are no adjustable parts or parts that can be replaced by the user within the equipment Taking out the Control Panel in order to have an access to the internal p...

Page 3: ...e 6 2 Control panel 7 3 Power supply 8 3 1 Battery condition 8 3 2 Battery charger 8 4 Measurement 10 5 Messages 12 6 Some notes about accuracy 13 7 RS232 output 14 8 Cleaning 14 9 Replacement fuse 14...

Page 4: ...d Digital reading alphanumeric display Up to 4 digits readings Powered by rechargeable battery or mains supply 0 1 resolution 200 maximum reading 1 1 Operating principle This device uses the Kelvin Br...

Page 5: ...ial terminal P Current terminal C RS232 data output Alphanumeric display Shows both the measured resistance value and messages to the operator Start key Stop key Hold key It retains the last measureme...

Page 6: ...rge level is low After a few minutes the measurement will be automatically interrupted in order to preserve the battery from a deep discharge that is prejudicial for its expected useful life 3 2 Batte...

Page 7: ...rt charging it as many times as is needed However the battery could be damaged if remains in deep discharge for a while To avoid this effect charge the battery before left the equipment in storage and...

Page 8: ...est current selector choose the range and the current to be used 3 Choose the power supply to use If you are going to use the battery advance to the following point If you are going to use the mains s...

Page 9: ...crease while test current decreases The equipment accuracy is specified for test currents higher than 80 8 The display will show the resistance value measured and the corresponding unit ohms m milli o...

Page 10: ...ars at the beginning of each test and it keeps on being there up to the operator rotates the test current control clockwise as necessary for the test current to be higher than 10 of the nominal curren...

Page 11: ...l offset Thus it is not necessary to carry out measurements by reversing the polarity in order to compute the average value Nevertheless if the operator suspect that there is a difference of temperatu...

Page 12: ...nce units are shown with the following symbols uR micro ohm mR milli ohm R ohm 8 Cleaning The panel terminals and connectors of the equipment must stay dry and clean Cleaning should be made using a we...

Page 13: ...tal direct reading of very low resistances in the alphanumerical display with up to 4 digits Very fast and accurate measurements Serial data output RS232 4800 bps Suitable for data collection in an ex...

Page 14: ...uding battery and accessories Dimensions 378 x 308 x 175 mm Accessories 2 Combined current and potential leads 1 8 m 1 Communication cable RS232 1 Power cord 1 User guide 1 Carrying bag Subject to tec...

Page 15: ...he failure was caused by misuse alteration accident or abnormal condition or handling you will be charged for the repair and the repaired equipment will be returned to you transportation prepaid This...

Page 16: ...18 Notes...

Reviews: