25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-70-
3
not valid for MBC600
4
not valid for Arco-1000
Mode
4
power output at
[
Ω
]
power range
[W]
nominal
frequency
[kHz]
output voltage *
[V
PP
]
Force 1
5W [200
Ω
]
1W 30W
**
860
Force 2
30W [200
Ω
]
5W 55W
**
1150
Force 3
55W [200
Ω
]
30W 80W
**
1300
Force 4
80W [200
Ω
]
55W 120W
**
1320
Force 5 ***
120W [200
Ω
]
80W 120W
**
1320
Spray 1
20W [200
Ω
]
20W 40W
**
9000/4800
Spray 2
40W [200
Ω
]
20W 60W
**
9000/4800
Spray 3
65W [200
Ω
]
40W 80W
**
9000/4700
Spray 4
90W [200
Ω
]
60W 100W
**
9000/5800
Spray 5
120W [200
Ω
]
80W 120W
**
9000/5800
Argon 1
1
20W [200
Ω
]
20W 40W
**
9000/4800
Argon 2
1
40W [200
Ω
]
20W 60W
**
9000/4800
Argon 3
1
65W [200
Ω
]
40W 80W
**
9000/4700
Argon 4
1
90W [200
Ω
]
60W 100W
**
9000/5800
Argon 5
1
120W [200
Ω
]
80W 120W
**
9000/5800
Argon 1
(2,5 probe & Argo-
Troc)
1
20W [200
Ω
]
20W 35W
**
9000/4800
Argon 2
(2,5 probe & Argo-
Troc)
1
35W [200
Ω
]
20W 50W
**
9000/4800
Argon 3
(2,5 probe & Argo-
Troc)
1
50W [200
Ω
]
35W 65W
**
9000/4700
Argon 4
(2,5 probe & Argo-
Troc)
1
70W [200
Ω
]
50W 80W
**
9000/5800
Argon 5
(2,5 probe & Argo-
Troc)
1
90W [200
Ω
]
65W 90W
**
9000/5800
Argon 1
(1,6 probe)
1
20W [200
Ω
]
20W 30W
**
9000/4800
Argon 2
(1,6 probe)
1
35W [200
Ω
]
20W 40W
**
9000/4800
Argon 3
(1,6 probe)
1
50W [200
Ω
]
30W 50W
**
9000/4700
Argon 4
(1,6 probe)
1
65W [200
Ω
]
40W 60W
**
9000/5800
Argon 5
(1,6 probe)
1
80W [200
Ω
]
50W 80W
**
9000/5800
Bipo. Cut
2,3,4
1 - 130W [200
Ω
]
-
-
360
330 - 760
Bipo. Coag
2,3,4
1 - 100W [100
Ω
]
-
-
360
160 - 460
*)
Output voltages are no-load values
**) pulse-shaped
signal
***)
ATTENTION: increased power only since software version 3.20. Older units
might have lower power output
Ignition/Working voltage