20. PROGRAMMABLE PARAMETERS
20.3.2 .
Therapy parameters in VT zone
ATP 1 program
Values
ATP program
OFF-Burst-
Burst+Scan
-Ramp-Ramp+Scan
Number of sequences
1-2-
3
-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-
8
-9-10-11-12-13-14-15
Cycles added per sequence
0
-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-
80
-85-90-95
Ramp decrement (per cycle) (ms)
0
-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-
8
-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-
2
-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-
220
-235-250-265-280-
295-310
ATP 2 program
Values
ATP program
OFF-Burst-Burst+Scan-
Ramp
-Ramp+Scan
Number of sequences
1-2-
3
-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-
6
-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-
1
-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-
85
-90-95
Ramp decrement (per cycle) (ms)
0-4-
8
-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0
-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-
2
-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-
220
-235-250-265-280-
295-310
Shock program
Values
Shock 1 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Shock 2 (J)
OFF
-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-
10-12-14-16-18-20-22-24-26-28-30-32-34-42
Number of Max. Shock (42 J)
OFF-1-2-3-
4
50
SORIN – INTENSIA VR 124 – U153A