13. DEVICE DESCRIPTION
─
Programmer-controlled VT induction sequences,
─
Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),
─
Programmable electrogram vectors (EGM A, EGM V, RVcoil-CAN, SVC-CAN, RVcoil-
SVC, LV bip, LV tip-RV),
─
Real-time annotations displayed with the markers and indicating the majority rhythm,
─
Manual ATP sequences,
─
Manual shocks.
─
Rescue shock
─
Follow-up tests:
─
Pacing lead impedance,
─
Coil impedance,
─
Capacitor charge time,
─
Sensitivity test,
─
Pacing threshold tests.
─
Data storage:
─
Therapy History Report,
─
Statistics (pace/sense, therapy, shocks, and battery voltage),
─
Up to 14 complete Holter records with event logs, marker channel notation, and
electrogram records.
The connector head has three ports:
─
Atrial “SonR” port: performs atrial bipolar pace/sense if a conventional IS-1 lead is
connected,
─
LV “IS-1” port: performs left ventricular bipolar pace,
─
RV “DF-4” port: performs right ventricular bipolar pace/sense, port for RV/SVC
defibrillation coils.
Distal lead terminal connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the device via
feedthroughs.
Programming System:
The Sorin programmer is used in conjunction with specific programmer software to
interrogate and program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring:
INTENSIA SonR CRT-D 184 is also equipped with the RF wireless technology which
enables to remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed
at home.
SORIN – INTENSIA SonR CRT-D 184 – U150A
31