2-8
(E)
MG40 Series
Connections
Establish the TCP connection from the soft ware.
1
Make the connection to the command interface. (Refer to
“2-2-1. Connection to the Command
2
Establish the connection settings given below.
Protocol TCP
IP address
Same as for the command interface
Port number
Port number designated by the command
3
Make the connection to the TCP server of the MG41 main unit.
No connections are required if UDP has been selected as the protocol. When UDP is selected, prepare to
receive at the designated port number.
2-5-2. Starting the Measurements
Switching over to the measurement mode
Aft er the settings of the measurement conditions have been completed, switch over to the measurement
mode.
1
Use the MOD command to set “1” which signifi es the measurement mode.
MOD=1 CR LF
OK000 CR LF
Starting the data transmission
Set transmission start using the data transmission control settings.
1
Use the NDT command to set “1” which signifi es transmission start.
NDT=1 CR LF
OK000 CR LF
When the data output control command is sent, the data is sent continuously from the data interface so the
necessary processing must be performed by the soft ware.
Th
e data sent is in the binary format. For details, refer to
.
Stopping the data transmission
Set transmission stop using the data transmission control settings.
1
Use the NDT command to set “0” which signifi es transmission stop.
NDT=0 CR LF
OK000 CR LF
Reference
Reference
Summary of Contents for MG40 Series
Page 2: ...MG40 Series ...
Page 4: ...ii J MG40 Series ...
Page 8: ...1 4 J MG40 Series ...
Page 20: ...3 4 J MG40 Series ...
Page 68: ...5 6 J MG40 Series ...
Page 72: ...ii E MG40 Series ...
Page 76: ...1 4 E MG40 Series ...
Page 88: ...3 4 E MG40 Series ...
Page 136: ...5 6 E MG40 Series ...