87
Overview
You can connect a commercially available touch panel
monitor or mouse to the unit, and use them to operate the
menu screens.
This chapter describes the settings for using a touch panel
monitor or mouse, and the screens that appear during
touch panel or mouse operations.
If you set one of [USB 3] to [USB 6] to [Touch Panel /
Mouse] in the [System Admin Settings] screen – [Device
Settings] screen – [Device 1] tab in the touch panel /
mouse mode while a compatible device is not connected,
operation will become impossible. Should you
inadvertently disable operation, you can restore the
standard operation mode by performing the following.
A
Press the MENU button on the front panel of the
unit.
A confirmation message for starting in standard mode
appears on the front panel display.
B
Select [OK], and press the ENTER button.
The unit restarts in standard mode.
• The minimum requirement for the touch panel monitor
resolution is 1366 × 768.
• Be aware of the following when setting one of [USB 3]
to [USB 6] to [Touch Panel / Mouse] in the [System
Admin Settings] screen – [Device Settings] screen –
[Device 1] tab.
– You must restart the unit after applying the setting.
– After applying the setting, the
V
,
v
,
B
,
b
, ENTER,
and MENU buttons on the front panel of the unit and
on the infrared remote control unit are disabled.
• For details on supported touch panel monitors, contact
your local Sony representative.
Configuring Touch Panel /
Mouse Settings
Configure the following settings to enable use of a touch
panel monitor or mouse.
1
Select [Device Settings] in the [System Admin
Settings] screen, and press the ENTER button.
For details on displaying the [System Admin
Settings] screen, see “Displaying the [System Admin
Settings] Screen” (page 64).
The [Device Settings] screen appears.
2
Display the [Device 1] tab, and set the USB port to
which the touch panel monitor or mouse will be
connected ([USB 3] to [USB 6]) to [Touch Panel /
Mouse].
3
Select [Advanced], and press the ENTER button.
The [Advanced] settings screen for touch panels and
mice appears.
4
Set [Display Cursor] to [Use], select [Apply], and
then press the ENTER button.
When [Use] is selected, a mouse cursor will appear
on the screen while using touch panel mode or a
mouse.
The default setting is [Not Used].
When you change this setting, a message indicating
that you must restart the unit appears.
5
Connect the touch panel monitor or mouse, and
follow the instructions on the screen and restart the
unit.