PAT-806-IT
– USER MANUAL
33
The measurement ends after a preset time runs out or
after pressing
STOP/ESC
.
After measuring is completed, read the result.
Correct result: R
ISO
> LIMIT
Incorrect result: R
ISO
< LIMIT
Notes:
- Before the measurement (also in AUTO test) check the resistance of the protective conductor R
PE
,
which should be correct.
4.5.2 R
ISO
measurement on devices of Class II (III)
Connect the mains plug of the tested device into the test socket of the me-
ter. L and N are shorted. Use the probe connected to socket
R
ISO
-
and touch
the conductive accessible parts of the tested device.