- 21 -
What to do?
If the recording was completed automatically, the device will switch itself off. If required, the recording
can be ended manually see chapter 4.2.6.
Use the black eject mechanism on the bottom of the device to remove the Compact Flash Card.
Never remove the card without using this button.
Insert the Compact Flash Card into the Compact Flash Card Reader.
Start the
DOMINO
software and select the Data Transfer
icon. The Recording Information will
be displayed. If required, enter a Description for this recording. If the
Auto-processing
checkbox is
selected, the recording will be downloaded and analysed automatically. To configure Auto-
Processing, click on the Configure Button. Here, it is possible to instruct the system to Analyse,
Display Report or Print the Report automatically after recording is downloaded.
Lastly, click on the
Transfer
button to download and process the recording.
4.2.9
Relevant information for Performing Measurements
The
SOMNOscreen™ plus provides several safety features. These will inform you of problems that
prevent a successful recording:
-
Insufficient capacity of the Rechargeable Battery
-
Compact Flash Card is Missing
-
Compact Flash Card Memory is insufficient for the programmed recording
-
A wrong montage was chosen
If any of these messages appear, use the following guide to correct the problem:
4.2.9.1 Insufficient Battery Capacity for the Programmed Duration
Display
Function
The capacity of the rechargeable battery is insufficient for the programmed
measurement. Select
MENU
on the display with S3
.
If
IGNORE
is selected with S3
, the system will recalculate the maximum
duration and display it.
OR:
OR:
If the Batteries capacity is less than 30%, a Recording will NOT be permitted.
The display will automatically change back to the Main Menu.
Switch off the device using the EXIT menu and confirm with S3
. Replace
the Battery with a fully charged one.
-
Open the Battery Compartment Cover and remove the battery. Insert it into the
Battery Charger.
Insert a fully charged battery into the device and start the recording again.
Please note:
When replacing the battery, only use batteries supplied by SOMNOmedics! If
alternative batteries are used, SOMNOmedics cannot guarantee error free
operation.