SET UP
Remove your current tubing from the top port of the
flowmeter if installing on existing system. If on a new
system connect black tubing from rear AMD top right
port (2) to top of flowmeter barb fitting(4).
Connect the silicone tubing from (1) port to vaporizer
inlet endcap Attach the inlet endcap onto the
vaporizer with a firm push and twist motion to secure
it.
1
2
3
4.
5.
1
2
3
Outlet
Inlet
4
If existing system remove tubing from vaporizer outlet
endcap and discard tubing. If new system connect
silicone tubing from (3) port to outlet endcap of
vaporizer. Attach the outlet endcap onto the vaporizer
with a firm push and twist motion to secure it.
Connect oxygen (carrier gas) supply source to the DISS
Male fitting on the back of the flowmeter.
Attach the color coded fresh gas delivery lines into the
3 common gas outlets on the front of the AMD-3.
5