Ad1:1st alarm mode setting;VL:low limit alarm of voltage
VH:high limit alarm of voltage,AL:low limit alarm of current
AH:high limit alarm of current
,
ex-factory setting is AH
SET
→
SET
>
3S
HY1:1st alarm hystersis setting;setting range-1999
~
9999
.
Eg:If alarm mode Ad1 be set to AH high limit alarm of current,
and AL1 set to 4.5,HY1 set to 0.1, when actual measured
current value
≤
(4.5-0.1), it will stop alarm
。
Note: other codes
’
alarm setting are similar to the above ,ex-factory setting is 0.100
bAd: baud rate , Baud rate option: 4.8Kbit
、
9.6Kbit , Ex-factory setting is 9.6Kbit,
SET
SET
SET
SET
→
SET
SET
SET
SET
SET
SET
SET
SET
SET
Ad2:Please refer to Ad1
,
AL2: Please refer to AL1
,
HY2:Please refer to HY1
,
ex-factory setting is AL
ex-factory setting is 0.500
ex-factory setting is 0.100
KKEH8E01-A/0-3
SET
SET
Measuring status
→
→
AL1:1st alarm setting menu ,setting range:-1999~9999 .
eg.If the first alarm mode Ad1 set to AH (high limit alarm
of current) and the AL1 set to 4.5 , when the measured
current value
≥
4.5 ,
the meter will alarm automatically .
Ex-factory value is 4.500
Ct: Current transform setting:
settable range is 1.0-1999, only for current channel. E.g.,
if the meter need to measure 20A current, the CT value
should be set as 4.000 (20/5=4.000), since the max.
current that can be input to the meter is 5A. Ex-factory
setting of CT is 1.00 .
Note: A CT (current transformer) should be equipped with
the meter if the measured current is higher than 5A.
LCK:Menu lock:setting value 000 or 010
when it is 000 , the parameters could be
modified , when it is 010 , all the parameters
couldn
’
t be modified. Ex-factory value is 000
APS:Current Amendment value . the range is
-1999 to 9999. It is to amend the measured
current, Display value=Measured value+APS
value, Ex-factory setting value is 0.000.
VPS:Voltage Amendment.value . the range is
-1999 to 9999. It is to amend the measured
voltage, Display value=Measured value+VPS
value, Ex-factory setting value is 0.000.
rL:Analog output low limit setting value,value
range is-1999 to 9999 . If it is set as current
analog output, the value is to be low limit of
measured current . Ex-factory setting is 0.000
rH:Analog output high limit setting, value
range is-1999 to 9999. If it is set as current
analog output, the value is to be high limit of
measured current . Ex-factory setting is 5.000
brm:Analog output mode selection is as below:
A: Current analog
V: Voltage analog
Ex-setting is A
Add:Commnunication address.
Settable range is 0-255.
Ex-factory setting is 000.
▲
2.Press SET key to select the parameter needing modified ,press to make the parameter flick ,and press to set needed value
and press SET key to confirm the modifying .Press to make the needing modified parameter flick, press
▲
+ to shift the decimal
point, press SET key to go on for another menu.
3.Under setting status
,
long press SET>3s can return to measuring status .
▲
▲
The instruction for the above menu operation and settings are as below:
1. Press SET key for 3 seconds
,
the meter will show the parameter setting menu.
/
▲