Printed in Canada
February 5, 2018
(#113872)
(Page 2 of 2)
For further information contact: Solinst Canada Ltd.
Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023
35 Todd Road, Georgetown, Ontario Canada L7G 4R8
Web Site: www.solinst.com E-mail: [email protected]
8. Use the 3/16" nut driver to undo the four nuts holding the controller
board assembly in place. Remove any washers. Be careful not to
lose any washers or nuts.
9. Slowly pull to remove the controller board assembly from the back
of the panel. This may take some force as the assembly may be
stuck to the adhesive on the back of the button faceplate label.
10. Leave any washers below the controller board assembly in place on
the mounting screws.
11. Peel any protective plastic from the LCD on the new controller
board assembly.
12. Place the new controller board assembly over the four mounting
screws on the back of the panel. Make sure the controller board
assembly is oriented correctly.
13. Put any washers back over the four screws. Use the nut driver
to tighten the for nuts, securing the controller board assembly in
place.
14. Reconnect the label ribbon cable to the circuit board.
15. Refasten the solenoid assembly to the panel and reconnect any
tubing by pushing into the fittings.
16. Plug the battery and solenoid connectors into the circuit board.
17. Reconnect any other tubing removed by pushing back into the
fittings.
18. Place the panel in the Control Unit case.
19. Reinstall the batteries. Test the Control Unit to ensure proper
connections were made.
20. Reinstall the four hex screws.