Troubleshooting & FAQs
29
THE BUS+ User Guide
Calibration Marks '8', '12', '16', '20'
When you are happy with the '4' mark calibration, repeat the same process to line up marks
'8', '12', '16' and '20' as closely as you can using the
LF FAST
and
HF FAST
switches. Use
the left-hand side
LOW THD
and
F/B
switches to go to the previous or next line-up positions,
should you want to make any changes.
Calibration Marks Indication
In addition to the Bus Compressor Meter Needle moving towards each mark, the following LEDs light red to indicate which mark
is currently being calibrated:
• EXT S/C D-EQ Left LED = 0 dB mark mechanical alignment
• Classic Stereo = 4 dB mark calibration
• Sum SC Stereo = 8 dB mark calibration
• Dual Mono = 12 dB mark calibration
• Mid Side = 16 dB mark calibration
• EXT S/C Comp Right = 20 dB mark calibration
Saving The Calibration
Important: To save the current calibration and exit, press and hold the
MODE
key and the unit will power cycle. If you simply power cycle the
unit from the mains switch, the calibration will
not
be saved!
Use the LOW THD and F/B
keys to move to the previous or
next calibration mark.
Press and hold the MODE key to save the calibration.