9
The calibration function will adjust the sensitivity according to the measured level and the
calibrator level at the calibrator frequency.
Click OK to accept.
Repeat for the second microphone.
Note: The calibration level can be checked by clicking “Check” instead of “Run”. This allows to
measure the calibration level without changing the sensitivity. It is recommended to check the
calibration after each measurement campaign to validate the correct operation of the probe.
Click on the “Enable Phase Comp.” button. It is recommended to enable the phase mismatch
compensation to increase the Pressure Residual Intensity Index and thus increase the dynamic
capability of the probe.