14
5.4 Check weighing
5. Operation
8
0
5
3
0
8
2
3
8
8
5
8
2
8
CAL
CAL0
LOAD
2000
5000
DONE
Place the reference object on the scale . Tare . Remove the reference object .
Place the object under test on the scale and read off the plus/minus difference
on the display.
5.5 Total weight display
When objects are on the scale, press the
key.
0
is displayed.
Remove the goods for weighing . The total weight appears with a minus sign .
8
0
5
3
0
8
2
3
8
8
5
8
2
8
CAL
CAL0
LOAD
2000
5000
DONE
5.6 Switching off
When the display shows
0
, press the
key. When operating the scale with
batteries or rechargeable batteries, the auto-off function is active . When the dis-
play shows
0
, the scale switches off automatically after 2 minutes.
8
0
5
3
0
8
2
3
8
8
5
8
2
8
CAL
CAL0
LOAD
2000
5000
DONE
5.7 Clear tare
When the scale is empty, press the
key again.
5.8 User calibration
Turn on the scale
, press and hold
key for one second then press
key for three times, to
enter calibration mode. At this moment, LCD display will show its Zero Counts.
Press
Key to zero the counts showed.
Then load the calibration weight according to its capacity (500 g or 5 kg).
Press
key again to calibrate the scale.
Wait till “CAL” disappears . Unload the calibration weight and test the scale .
Summary of Contents for 9202