Measuring Capacitance
WARNING!
All capacitance measurements are to be made on de-
energized circuits with all capacitors discharged only.
Failure to de-energize and discharge capacitors prior
to measuring them could result in instrument damage
and/or personal injury. Make sure the temperature
probe is NOT plugged in during this test.
Instrument set-up:
FUNC.
BLACK
RED
TEST LEAD
TEST LEAD
COM
V/
W
/
m
A
Measurement Procedure:
1.
Disconnect power to circuit to be measured.
2.
Plug black test lead into the COM input jack.
3.
Plug red test lead into V
/
W
input jack.
4.
Set the rotary switch to the position.
5.
Remove the capacitor from the circuit and dis-
charge it.
6.
Connect test leads to the capacitor to be measured.
Observe polarity on polarity sensitive capacitors.
7.
Read the capacitor value on the LCD.
Optional Modes
(See Other Features Section)
•
D-H:
Freezes the reading on the LCD.
21