background image

Setting the Calibration Modes 

 

Automatic Calibration

This mode is useful  when  searching for leaks normally  and also  in 
heavily contaminated areas. The detector automatically recalibrates 
(re-zeroes) to ambient  levels, which stops the constant alarm, only 
responding to higher levels of refrigerant. This allows searching for 
the  true  source  of  the  leak.  The  detector  will  reset  automatically 
after 3 or 4 seconds and stop the detector from alarming.  Note: the 
detector  automatically  defaults  to  this  mode  when  the  unit  is 
powered on normally.  

 

Manual Calibration: 

This mode may be useful when more control over the calibration is 
required  and  also  when  searching  for  smaller  leaks.  In  this  mode 
the  detector  recalibrates  to  ambient  levels,  but  only  when  you 
choose  to  do  so.  The  calibration  will  remain  constant  when 
searching for small leaks.  

 

To  use the  detector  in  Manual  Calibration  mode:  Press and  hold  the  SEN 
button  and  release  the  button  when  the  SEN  LED  level  indicator  starts  to 
blink  or  (flash).  Note:  after  resetting,  the  LED  will  blink  rapidly  until  the 
manual reset cycle is complete. 

 

To  return  to  Automatic  Calibration  press  and  hold  the  SEN  button  and 
release the button when the SEN LED stops blinking (flashing). 

 

Note:  the  sensitivity  levels  can  only  be  changed  in  Automatic 
Calibration  mode.  To  change  sensitivity  levels  while  in  Manual 
mode, return to Automatic mode, select the desired level and return 
back to Manual mode. 

 

 

Low Battery Indicator 

 

Replace  the  4  AA  Alkaline  batteries  when  the  red  LED  on  the 
control  panel  is  lit.    Follow  battery  installation  instructions  under 

Maintenance

 section. 

 

 

Audio Mute Function 

 
To  silence  or  mute  the  audio  beep  and  alarm  signal,  press  the 
MUTE button.   To restore the audio sound, press the MUTE button 
again. (Note: a few seconds is required to restore sound if the mute 
button is pressed in rapid succession.) 

 

Adjusting Sensitivity Levels 

 

The  Leak  Detector  will  default  to  the  NORM  sensitivity  level 
automatically once the unit comes out of the warm up cycle and 
the green LED will turn on.  
 
To  change  sensitivity  levels,  press  the  SENS  once  for  HI 
sensitivity  (red  LED  will  turn  on)  and  again  for  LO  sensitivity 
(yellow LED will turn on). 

 

Leak Test Vial 

 

The leak detector comes with a Leak Test Vial that allows the 
user to make sure the detector is performing properly.  To test: 
 

1.  Remove the Plastic Seal Cap on the top of the Leak Test 

Vial  by  pulling  it  off  (see  fig.  below).  Also,  remove  and 
discard the circular Leak Test Cap Seal.  

2.  Turn  on  the  leak  detector  and  allow  the  instrument  to 

complete the warm up cycle.  

3.  Place the sensor close to the small hole in the top of the 

Leak  Test  Vial.  The  beep  rate  should  increase  and  the 
Digital Leak Size Indicator should display a number from 
4-6  indicating  that  the  sensor  and  the  electronics  are 
working properly.  

 

 

 

 
 
 
 
 
 

 
 
 
NOTE:  Remember  to  replace  the  Plastic  Seal  Cap  after  the  leak 
test  is  completed.  Replace  the  Leak  Test  Vial  when  the  green 
media color is no longer visible and the leak media has evaporated. 
Remove  the  threaded  cap  on  the  vial  to  confirm  the  status  of  the 
leak  test  media.  After  the  leak  test  media  has  evaporated,  it  is 
normal for there to be a green film on the wall of the Leak Test Vial.  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Remove the Plastic  
Seal Cap to Test the 
Leak Detector 

Leak Test Cap Seal 

Leak Test Vial 

Summary of Contents for ACT760B

Page 1: ...stry first two year warranty on the sensor This warranty applies to all repairable instruments that have not been tampered with or damaged through improper use including unauthorized opening of the unit Please ship warranty units that require repair freight prepaid to Service Center along with proof of purchase return address phone number and or email address Snap on Wrench S and There is a Differ...

Page 2: ...ds Dextron Transmission fluid heated to 160 ºF N N A Quaker State Motor Oil heated to 160 ºF N N A Rain X Windshield Wash Fluid N N A Ford silicone lubricant N N A Ford Rust Inhibitor when wet Y Yes Ford Gasket Adhesive when wet Y Yes Loctite Natural Blue degreaser undiluted Y Yes Ford Brake Parts Cleaner when wet Y Yes Ford Silicone Rubber when uncured Y Yes Motorcraft Antifreeze heated to 160 ºF...

Page 3: ...tionally if the instrument becomes unstable during its operation it is an indication that the sensor may be defective or dislodged 5 2 ACT760B Control Panel Operating Instructions 2 TURN ON Press the ON OFF button once to turn on and again to turn off 3 WARM UP The detector automatically starts heating the sensor During the heating cycle the detector will sound a slow beep Warm up is usually less ...

Page 4: ... battery installation instructions under Maintenance section Audio Mute Function To silence or mute the audio beep and alarm signal press the MUTE button To restore the audio sound press the MUTE button again Note a few seconds is required to restore sound if the mute button is pressed in rapid succession 3 Adjusting Sensitivity Levels The Leak Detector will default to the NORM sensitivity level a...

Reviews: