6
appropriate position when it is out of the factory. When using
,
place the lamp
Head at 1m away from the surgical site. If it needs to adjust slightly, rotate the handle
in order to the site which meet the surgical needs.
Handle’s remove
When disinfecting, hold the lower part of the handle with hand, and press the
button with your thumb, pull down. Then you can remove it. When installation, keep
the holes on the direction of the button, and push it gently until the button pops out.
Handle’s disinfection
Do not put the handle in high pressure steam sterilizer, in order to avoid
deforming. It can use peroxyacetic acid solution, glutaruc dialdehyde solution and
medical alcohol to disinfect in normal temperature.
7. Maintain
Damping adjustable rotating arm
The center seat of 2 rotating arms has 2 damping adjustment screws. If it drifts
after using for some time, you can tighten the damp with wrench according to picture
8.
picture 9
Lamp head lamp adjustment
Lamp head and two brackets should have appropriate damping,to avoid
transposition. If appearing the phenomenon, tighten the screws properly according to
picture 9.
Balance arm adjustment
Please operate it according to the instruction of Ondal.